The new name for "Hansens"
What is Dialyzer Connectors
Color of buttons that are selectable, allowing changes to be made
What is blue
This replaces normal saline on the 5008x machine
What is Substitution Fluid
Location to check for residual bleach
What is wall drain
External cleaning of the machine requires this solution using a damp (NOT saturated) cloth
What is 1:100 bleach solution
This takes place of the Arterial Blood Chamber
What is the Arterial Pressure Dome
Choose this menu for viewing the machine set-up instructions
What is Blood System
The amount of substitution fluid the machine will prime with
What is 500mL
If using a bibag, this test is no longer required for the 5008x
This is completed every 7 days
What is cold disinfect (bleach)
These doors are easily removable and replaced
What is EBM doors (extracorporeal blood circuit module doors)
Go to this menu option prior to connecting to patient, in order to stop blood pump and exit priming
What is Preparation
This piece provides the substitution fluid
What is Substitution Fluid Port
Name the 3 components of dialysate
RO, acid, bicarb
This is completed at the end of every day
What is chemical heat disinfect
These require an exchange every 90 days, 100 heat cycles, OR 13 cold disinfects (bleach) - whichever comes first
What is the Diasafe Filters
OCM, AFM, BPM, Circulation - are all found under which menu
What is Options Menu
This creates a sterile, non-pyrogenic substitution fluid
What is Diasafe Filter #2
This is an independent test performed by the 5008x, that tests for electrical currents of dialysate
What is conductivity
This is required when contamination of either port is suspected
What is 51 minute mandatory disinfection
This must be used with 60cc syringe to open the Arterial Pressure Measurement Unit after a power outage
What is the Check Valve
When delivering a substitution fluid bolus or changing tx modality, we would choose this menu
What is Online
List the 6 components that make up the Online System
What is Diasafe Filter #2, Substitution Fluid, Substitution Fluid Pump, Substitution Fluid Port, Rinse Port, Safeline
This indicates that dialysate is on and flowing
What is the green waves with white moving bar
Citric Acid is being replaced with this premixed solution
What is CitraSure