The filters on the chemistry analyzer are recommended to be cleaned at this frequency.
What is monthly?
These are the only types of tubes that can be used to run a CBC.
What is an EDTA tube?
This sample type can be used when running an ePOC card and COAG+ strip.
What is untreated whole blood?
This is how much urine sample volume is needed for the AIM cartridge.
What is 0.55 mL?
This is the total number of accounts you need to visit weekly in order to leave a section on your WAR blank.
What is 12?
RE fluid is good for this many days after opening.
What is 90 days?
The HT5+ has this many reagents.
What is 7?
This is the software version the Ei+ is on currently.
What is 2.22.0?
During installation, these 2 analyzers have zip ties that need to be cut and removed prior to turning them on.
What is the HT5+ and Slide Scanner?
Accuplex can confidently detect this many canine vector-borne pathogens.
What is 7?
These are two common sample types that can affect chemistry results.
What are hemolytic and icteric samples?
When a clinic has run out of probe cleanser for the HT5/HT5+, this reagent can be used as a short term solution.
What is Diluent (DS Diluent for HT5+)?
The COAG+ can effectively run a PTT/aPTT on a patient with a PCV within this range.
What is 20-70%?
This is the maximum amount of images the AIM displays after completing a full scan.
What is 40 images?
Virtual trainings and installs on your WAR should be listed in what color font?
What is purple?
It takes approximately 11 minutes to run this chemistry panel.
What is a Comprehensive Plus Panel?
Underfilling EDTA tubes can cause this to happen to RBCs.
What is shrink?
This test can be ran on the Ei but not on the Ei+.
What is Bile Acids?
When using the HVSS/Slide Scanner, this slide should not be scanned and submitted.
What is an ear cytology?
When installing with a cloud based software, this item is required to be inputted into the DCU.
What is a gateway string?
This is how many uL (microliters) of sample are used per slide.
What is 10 uL, 15 for electrolyte?
This is the maximum of amount of time a purple top tube can be on a rocker to keep the best sample integrity.
What is 30 minutes?
T/F - ePOC cards can be used after being refrigerated.
What is true?
Need to bring them back to room temperature and can be used for 8 days.
This item is discarded when installing a Slide Scanner.
What is the Motic product manual?
These 2 codes can be submitted within 30 days of the original test submission to our reference lab.
What is monitoring and completion codes?