This occurs when the pump does not have sufficient preload.
What is PI event?
The MOMENTUM 3 trial measured superiority in centrifugal flow pump vs _____.
What is axial?
Patients should carry this extra equipment with them at all times.
What is the 2 fully charged batteries, clips and backup controller?
Target blood pressure is.
What is <90mmHg?
Power cable disconnected advisory alarm can be silenced for this number of minutes.
What is 2 minutes?
This occurs when there is a PI event?
What is drop speed by 200-400rpm?
This is the recommended daily routine for HeartMate 3 LVAD patients to ensure that their equipment is functioning properly.
What is checking the System Controller display (self test), checking the power sources, inspecting the driveline exit site, and ensuring they have their backup equipment nearby?
It is most important to avoid this to prevent driveline infections.
What is trauma?
Clinicians will make this change to the speed with an elevated PI.
What is none? Evaluate your patient and make clinical adjustments if necessary.
For 24 hour HM3 support, call this number____.
What is 1-800-456-1477 (Heartline)?
These 3 values are part of the calculated flow algorithm.
What is speed, power and hematocrit?
This is the specific mechanism by which the HeartMate 3 LVAD reduces the risk of pump thrombosis that cannot be turned off or adjusted.
What is artificial pulse?
This is the first step in order to perform a controller exchange.
What is power up backup controller?
List 3 potential causes for PI event.
Bearing down
Ventricular arrhythmias
Coughing/sneezi
Low preload (dehydration, bleeding)
Tamponade
Afterload decreases
RV dysfunction
Speed too high
Your patient is having "power cable disconnect" alarms. The log file captured intermittent indications of a weak connection between the batteries & clips. You should do this____.
What is check the battery and battery clip contacts for integrity and clean the battery & battery clip contacts with an alcohol wipe?
These 2 parameters trend synchronously.
What is flow and power?
The Momentum 3 5-year data revealed the following result for survival.
What is 58.4%?
You should be on this screen on the HM touch UNLESS you are in the OR.
What is "monitor"?
A low flow alarm could be the result of these 4 conditions.
Low preload
RV failure
Arrhythmia
HTN
There is a power outage while the patient is on the MPU. The____ will provide power until the patient goes onto batteries.
What is 11 volt lithium-ion emergency backup battery (EBB)?
These 3 features on ECHO are assessed in order to optimize speed.
What is aortic valve opening, left ventricular size, position of the septum? (p. 4-23, IFU)
This is the technology that HeartMate 3 LVAD uses to eliminate mechanical bearings and minimize blood trauma.
What is Full MagLev technology?
This should be avoided when a HM3 patient is travelling through the airport.
What is body scanners and metal detectors may interfere with your pump and should be avoided?
With hypertension, you would see ______ flow, _______ power, _____ PI.
What is:
Low flow
Low power
High PI
This is the only alarm that cannot be silenced in the HM3.
What is "system controller hardware fault" hazard alarm?