Decontam 1
Decontam 2
Decontam 3
Decontam 4
Decontam 5
100

This pressure relationship must be maintained in the decontamination area.

Negative air pressure.

100

Decontamination should have this number of air exchanges every hour.

10 air exchanges per hour.

100

The required temperature range in decontamination is between these two temperatures.

 18°C to 20°C

100

Relative humidity must remain within this required range, although 40% to 50% is preferred.

30% to 60%.

100

MDRT personnel must change this protective item whenever it becomes wet or dirty in decontamination.

PPE

200

These gloves should have extended cuffs that are pulled over the cuffs of the gown.

Nitrile gloves.

200

After discovering that a glove is torn, a technician must replace it and perform this procedure.

Hand hygiene.

200

An IV pole belongs to this Spaulding classification.

Non-critical.

200

A towel used to keep instruments moist during transport should be moistened with this liquid, never saline.

Water.

200

Preparation for decontamination begins at this location.

Point of use.

300

This immunization is identified as part of preventing blood and body fluid exposure.

Hepatitis B immunization.
300

These device-specific directions explain how a reusable medical device must be cleaned and sterilized.

Manufacturer's instructions for use (MIFU)

300

This is the most important step in medical-device reprocessing in decontamination.

Manual cleaning.

300

A flexible endoscope normally belongs to this Spaulding classification.

Semi-critical.

300

Instruments with box locks should be in this position while being decontaminated.

Open.

400

This emergency equipment must be easily accessible and tested regularly for exposures involving the eyes or mucous membranes.

Eyewash station and emergency shower.

400

Before a new device is used on a patient, the manufacturer must provide this education to MDRA staff.

In-service education/training.

400

Surgical instruments belong to this classification because they enter sterile tissue, the vascular system or a sterile body space.

Critical.

400

This is the minimum level of disinfection required for semi-critical devices that cannot be sterilized.

High-level disinfection (HLD).

400

Medical devices should be taken apart to these components before cleaning.

Simplest components.

500

Forced air used to dry channels and lumens should be filtered to this size.

0.01 micron

500

New powered medical devices must undergo cleaning and inspection before sterilization because manufacturers may apply these protective substances during shipping.

Oils and greases.

500

Soiled transport routes should avoid high-traffic areas and areas where these supplies are stored.

Sterile supplies.

500

A semi-critical medical device capable of withstanding sterilization must receive this form of processing instead of high-level disinfection.

Sterilization.

500

Ideally, cleaning should begin within this period following use.

15 minutes to 1 hour.