Feature Presentation
Put me in, Coach!
Good to Know...
Abort Mission!
Stretched Too Thin
100

No green light on your detachment handle? No problem! Embold coils don't need a detachment handle. Instead, they offer easy, handle-free detachment. Snap the pusher wire between the white perforation markers and pull back the proximal end this much to deploy the coil.

At least 1cm


100

This coil is the workhorse coil within the Embold family. Use it in small vessels, large vessels, high-flow & low-flow vessels, as a framing coil or as a backstop for subsequent coils!

Embold Fibered

100

Prep work to get Embold coils ready to deploy is minimal… but always do this prior to inserting your first coil, after using contrast, and after every couple coils to keep things moving smoothly.

Flush the microcatheter

100

Always be sure to do this to visually confirm that the coil has detached prior to retracting the delivery system. Inadvertent coil pullback can be a pain!

View with fluoro

100

These are some reasons why coil stretch occurs.

Multiple retractions/manipulations

   - incorrect sizing

   - coil not forming correctly 

Inadvertent coil pullback

   - couplers in MC

   - couplers in small vessel <1.5mm

Coil advanced or withdrawn against high resistance

   - intertwined coil nest, MC tip kept in coil nest while advancing

   - Pinning coil with MC

   - kink in MC


200

To reduce coiling complexity through fewer catheter exchanges, all Embold coils have multi-catheter compatibility. This means any Embold coil is compatible with any microcatheter that has an ID within this range.

.021" - .027" ID



200

I have the ability to fill up space and create a dense mechanical occlusion. I'm great in aneurysms or large vessels… but never use me on my own! I always need a dance partner.

Embold Packing

200

Radial entry is your access of choice for the embolization at hand. No problem! … unless you choose a microcatheter of this length.

175 cm


Embold can not be deployed through 175cm radial length microcatheter

200

Full retractibility is one of Embold's claims to fame. Keep at least one of these on the sterile table to protect the coil in case one is retracted and you plan to try using the coil again.

Introducer sheath


Bathe coil in hep saline once retracted & re-sheathed

200

Stretch can happen with any retractable coil. If it happens, early recognition of stretch is important to best manage the affected coil thereafter. On fluoro, this is the first telltale visual sign that coil stretch has or is beginning to occur.

Loss of coil radiopacity


300

Your microcatheter tip is buried in a coil mass. You can't see the end of it on fluoro, so you can't tell if your Embold coil has completely exited the microcatheter. This additional indicator will help you determine if the coil is advanced far enough.

RO Marker 3cm from proximal coil couplers


300

I'm best served in small vessels, near a vessel ostium, or in a small landing zone. A smooth, easy delivery with limited catheter kickback is my claim to fame.

Embold Soft

300

This coil has the most outward radial force, making it ideal for anchoring. This coil also has outward force, abeit 10x less than the former. Regardless, both can be used for anchoring and finishing in the right situations.

Embold Fibered = defined helical shape c/ anchorable wind, most radial force

Embold Soft = defined helical shape c/ softer wind, has radial force, but less

(Embold Packing = no outward radial force, oscillating shape, no anchorability, requires backstop)

300

Slow is smooth, smooth is fast. If your microcatheter is advanced very close to the coil mass, doing this with the MC prior to coil retraction will help ensure your Embold coil isn't withdrawn along a tight, angled pathway, and therefore will retract smoothly.

Pull back the Microcathether to improve retraction angle.

Coil should be in-line with MC when retracting.

300

From a tactile standpoint, coil stretch can present as a spectrum depending on the severity of the stretch. Minor stretch might present as a subtle "gritty" feeling, increased friction, a loss of 1:1 response, or a delay in forward motion & retraction. With significant stretch, you'll probably lose this ability altogether.

Loss of ability to push & pull

400

Any "ordinary" coil can't be advanced if its pusher wire gets one of these sharp twists in it. Good thing all Embold coils have a special nitinol delivery system that prevents this from happening.

Delivery wire kink

Nitinol delivery wire offers kink-less performance!


Bonus: what are the constituent metals of nitinol?

400

This coil does a fantastic job of filling up large spaces, achieving a dense coil mass. However, for best results, consider it for vessels that are 5-6mm or larger. For smaller vessels, consider Soft instead.

Embold Packing


400

This Embold feature, found between the coil and the nitinol delivery wire, promotes trackability and easy navigation through tortuous vessels.

25.5 cm Polymer Shaft


400

Coil not forming correctly, coil isn't sticking in place, coil is pulsating with systole under fluoro... these are all signs that the coil is sized in this manner and that it may need to be retracted.

Sized incorrectly. Ie. too big or too small

400

You've been working a coil quite a bit and you've identified that it might be starting to stretch. Most of your coil is already deployed and you're happy with its position. Taking this next step is probably the best path forward.

Advance coil to acceptable position & deploy completely

500

Embold Fibered coils have PET in the form of fibers to promote thrombogenicity. Embold Soft & Packing don't have fibers, but do have a PET suture inside the coil to combat this from happening to the coil. Luckily, it promotes thrombosis here, too!

Coil Stretch

Stretch-resistant PET suture runs from proximal to distal end of coil.

500

In small vessels (~2-5mm), size Soft & Fibered coils at this ratio. Vessel over 5mm? Try sizing them this way instead.

1:1 in small vessels up to ~5mm

Oversize coil 25% in larger vessels

(upsize to next coil size)

500

This Embold coil uses the least metal, having the thinnest primary wind but has a hairy occlusive helper. This coil puts the most metal in a vessel, using the largest primary wind. This coil falls in the middle.

Embold Fibered: Thinnest primary wind, least metal volume but fibered

Embold Soft: Intermediate primary wind thickness, intermediate metal volume

Embold Packing: Thickest primary wind, most metal volume

500

This is the most common cause for the coil couplers not separating when the detachment mechanism is actuated… and the coil inadvertently retracts along with the delivery wire.

Couplers constrained by external forces


Examples: Couplers inside Microcatheter, couplers buried within dense coil pack, couplers in vessel <1.5mm

500

An aggressive pullback has caused your coil to stretch significantly; you can no longer push/pull the coil but most of it is still within the microcatheter. Do this to save the day.

1. Don't panic

2. Don't detach coil

3. Bail out - remove entire system as a unit

   - Pin the pusher wire to the MC hub, pull the MC and the coil out together.


MC retraction is a nuisance. Foreign body retrieval is a disaster.