Quality
Documentation
Tests/Equipment/maintenance
Training
Work Instructions
100
What it means : QMS?

Quality management system

100

Primary tool used by lab to assist compliance with the QMS?

LIMS

100

WI applicable to Calibration?

WI0300

100

How can you demonstrate the training? 

Training Evidence: training report, quiz in LIMS

100

Procedures revision types and describe them:

>Editorial changes

>Procedure Revision

200

Mention the differences between Not conforming test and not conforming results? 

NCT: This may result in invalid test, the laboratory perform CMR correction per WI0101- documented on QCR.NCR: Test results that do not conform to the conformance requirements documented in the CMR shall have a Disposition Statement indicating lack of conformance.

200

For quality there are two types of documents:

1.-Documented information that is maintained (Procedures per WI0100) 2.-Documented information that is retained (Records per WI7300)

200

Whats is the Uncertainty  

Non-negative parameter characterizing the dispersion of the quantity values being attributed to a measured

200

What should include a training report?

Type of training approach used, training material used, what was done and the results, restrictions limitations, additional comments

200

you want a revision for a procedure, what you do?

change request in LIMS

300

All labs had quality audits: Nadcap aproved labs and non Nadcap  Labs, mention at least two of each one

Nadcap:Site quality audits; Periodic observation of test; Self audit; Network audit; Nadcap audit

Not Nadcap: Internal audits; Site quality audits

300

What you need to review in a CMR Results?

>CMR results are documented appropriately > transfer CMRs are completed.>result disposition appropriate for the Detailed Instructions and/or Conformance Requirements.>all SOP fields contain an appropriate and correct entry> CMR shall either be placed “on hold” for customer and external reviews or placed “in review” in accordance with Second Level Employee Review.>If the review of test results requires approval, as defined by CMR instructions or technical work instructions, the following statement shall be used: “I approve these test results”.

300

Mentions the differences between:Invalid Test; Replacement Test; Retest

•INVALID TEST: A test whose results are found to be unrepresentative based on identified causes other than the proprieties of the material being tested (e.g. errors in specimen machining or testing). When an invalid test is replaced, tests are recorded and cross referenced with the original tests, including explanations where applicable.

•REPLACEMENT TEST: a repeat of a test by the same laboratory, using the same method, equipment (of equivalent accuracy or better) and sample (same material, lot/heat, location, orientation as the original or on the original specimen if not destroyed). REPLACEMENT TESTS are performed in response to INVALID TEST.

•RETEST: A repeat of a TEST by the same laboratory, using the same method, equipment (of equivalent accuracy or better), and sample (same material, lot/heat, location, orientation as the original or on the original specimen if not destroyed). RETEST is performed in response to non-conforming test results and only when allowed by material or Honeywell specification, and are performed in accordance with the material or Honeywell specification

300

Which are the training Levels for a procedure?

Awareness, trained, certify, SME

300

What are the two types of procedures?

>Based in standard

>Based in Methods

400

Mention the WI applicable to Lab Quality

-WI0110: Quality Systems and Practices-WI8280: Quality System Compliance Program-WI0100: Work Instruction System-WI0101: CMR Work Flow Process-WI0300: Calibration of Equipment-WI0700: Equipment Maintenance-WI0106: Round Robin Testing-WI7300: Record Retention Schedule

400

Revision of released CMRS:

Edit CMRs, Activate Released CMRs

400

The laboratory uses a calibration schedule with frequencies in accordance with what?

Manufacturers’ recommendations or the individual checklist requirements for all internal and external calibrations performed on testing equipment. This schedule documents duration between calibrations (frequency) and responsibility for calibration (source).

400
Who perform the training in a procedure?

Process owner

400

What are the lab procedures auditables and the non auditables?

Auditables: Internal Calibration instructions; Lab procedures;Test Methods

Non Auditables:Standar operation Sheets (SOS); Equipment information; Lab Guidelines

500

What to do when transcription errors are identified in a released CMR?

proceed to activate CMR and correct error and document actions per WI0101 Transcription error correction may result in an invalid test or changes in disposition, in this case a RPS (rapid problem solving) may be necessary to avoid reoccurrence, as well as customer notification.

500

What to do when LIMS is down? and where to find it?

WI 5501, what to do when LIMS is down procedure

500

Which are the procedure for all equipment calibration?

1.-Define, establish and document calibration requirements 2.-Request/ perform calibration 3.-Obtain/ create calibration certificate 4.-Equipment Labeling 5.-Out of Tolerance(IF applicable)

500

Who can perform a Round Robin or PT?

All the analysts certified in the process

500

where can i find the procedure for WI publication?

WI0100

M
e
n
u