The Swiss Miss
Under a Microscope
Can You See Me?
What's Wrong Now?
Flash Star!
400

The number of motors or pumps that can be connected to a BB Board.

What are 2 motors or Pumps?

400

IFA is short for this.

What is Indirect Fluorescent Antibody?

400

The proper orientation of Slides in the Slide Carrier.

What is Barcode away from rubber piece?

400

This is likely the source of excessive scan times and "No Image" wells on a Nova View.

What is Z Cal?

400

You'll be waiting this long for a Bioflash to give you results.

What is 30 min plus one minute for every sample after.

800

The 2 files that are edited for Washer alignments on the 160/240.

What are Washer.ini & siasconfig.ini ?

800

These consumables labeled HEp 2-12 W, ANCA Ethanol and ANCA Formalin are required to run IFA assays.

What are Slides?

800

This should always be set to one.  

What is Coverslip correction Ring?

800

If you see a Reader 35 or 36 error on a Bioflash you should be looking at this module.

What is the Lumo / Lumo door?

800

what is the temperature Ranges for the incubator and Reagent Carousel?

1000

These 4 programs are used during a QL3K PM.  

What is X-UTIL, XAT, Scaling Cale, QL software  ?

1000

When the wash is properly aligned and configured on a QL3K, the bubble on the slides can be seen doing this.

What is nothing / shaking?

1000

These are the 3 calibrations completed on every PM and service Call.

What are XY Cal, Z Cal, FITC Cal?

1000

When these wear out, you’ll see an Unknown Module Error 00/08 on a QL3k.

What are LLD/ LQS ribbon cables?

1000

Either of these components may cause Trigger level errors in the Lumo after PM.  

What are the Lumo aspirator tubing / Coax cable?

1500

When creating an Assay for testing in the QL3K Software, this step is required before you can save the Assay.

What is the Distribution step?

1500

Blank wells, or stripes within wells that are blank, likely mean this height is too low.

What is wash height?

1500

The 3 components between the light source and the slide. (Module names)

What are light guides, collimator and scope?

1500

Most likely cause of abnormal reaction errors on BF results.

What is system contamination?

1500

Cuvette loader 2C errors are typically caused by this.

What is Shutter failed to close// stack stuck at top of chute?

2000

The number of Contrad to DIH20 to make system Liquid

What is 4ml to 1L? 

Or, .04% Contrad

Or, 24ML to 6L

2000

When tissue on a slide curls at the edges it's usually one or more of these causes.  

What are wash height, wash pump power, final aspirate?

2000

Incorrect Time and Date, PC not turning on, and Drivers not working are all signs it's time to replace this component.  

What is the CMOS Battery?

2000

When a QL3K thinks a full-wash buffer bottle is empty this component probably got wet and stopped working.

What is the CQ28 level sensor on Liquid station?

2000

These 4 files should be checked for size during a PM.

What are InstrumentLog.TAB QCResults.TAB ResultArchive.TAB Results.TAB ?