TVP Modes & Settings
Troubleshooting & Malfunctions
Nurse Responsibilities
Indications and Complications
Equipment and Setup
100

This commonly used pacing mode means the ventricle is paced, sensed, and pacing is inhibited if intrinsic activity is detected.

What is VVI?

Scenario: Patient with complete heart block and no atrial activity. 

Paces and senses the ventricle. Inhibits pacing if intrinsic ventricular activity is detected.

100

This issue is identified by pacing spikes without QRS complexes.

What is failure to capture?
Example: Pacing spikes without QRS complexes.

You check the ventricular output and it's set to 5 mA. You increase it to 10 mA, and capture is restored.

100

This must be done before making any changes to pacemaker settings.

What is consult the MRP?

100

TVP is indicated for this type of arrhythmia that causes unstable bradycardia.

What are heart blocks?

100

This device delivers electrical impulses to the heart.

What is the pulse generator?

200

In the DDD mode, this letter indicates dual chamber pacing and sensing.

What is D?

Scenario: Post-op cardiac surgery patient with intermittent AV block.

Dual chamber pacing and sensing. Responds to sensed activity by triggering or inhibiting pacing.  

200

This occurs when the pacemaker fires despite intrinsic activity.

What is failure to sense?

sensitivity setting set too high (e.g., 5.0 mV). You lower it to 2.0 mV, allowing the pacemaker to detect intrinsic activity and inhibit pacing appropriately.

200

This should be infused at a minimum of 10 mL/hr through the side arm.

What is normal saline?

200

This complication involves fluid accumulation in the pericardial sac.

What is cardiac tamponade?

200

This wire is inserted through a sheath and sits in the right ventricle.

What is the transvenous pacing wire?

300

This pacing mode does not respond to sensed activity and fires at a fixed rate.

What is DOO?

Asynchronous pacing of both chambers. No sensing; fires regardless of intrinsic activity.

300

A pacemaker that doesn’t fire at all may be experiencing this.

What is failure to pace?

Check generator, replace battery

300

This bedside item should be taped to the generator for emergencies.

What are spare batteries?

300

TVP may be used as a bridge to this permanent intervention.

What is permanent pacemaker insertion?

300

This part of the setup connects the pacing wire to the generator.

What is the bridging cable?

400

This setting determines how much voltage is needed for the pacemaker to detect intrinsic activity.

What is sensitivity (measured in mV)?

400

Increasing this setting may help correct failure to capture.

What is output (mA)?

400

This dressing allows visibility of the insertion site.

What is a transparent dressing (e.g., Tegaderm)?

400

This complication can occur if the pacing wire perforates the ventricle.

What is ventricular perforation?

400

This test is ordered after pacemaker insertion to confirm wire placement.

What is a chest X-ray?
Example: Ensures proper wire positioning.

500

This pacing mode is only appropriate when the patient has intact AV conduction. It paces and senses the atrium, inhibiting pacing when intrinsic atrial activity is detected. It should not be used in patients with AV block due to the risk of ventricular asystole.

What is AAI (Atrial Demand Pacing)?

500

This term describes when the pacemaker misinterprets noise as intrinsic activity.

What is oversensing?

Interpreting muscle tremors as intrinsic activity.  Sensitivity set to 0.5, increase to 2.0 reducing oversensing

500

This safety check is performed at every shift change.

What is RN-to-RN safety check?

500

This condition may result from air entering the bloodstream during insertion.

What is air embolism?

500

This measurement is taken from the black line to the hub of the Cordis.

What is the internal wire length?

M
e
n
u