This guideline outlines the standards to be used when a patient is suspected of having electrocerebral inactivity

What is ACNS Guideline 6: Minimum Technical Standards for EEG Recording in Suspected Cerebral Death

100

This header explains the imepances an ECI recording must have.  (Give ACNS Guideline number and full paragraph header)

What is ACNS Guideline 6: 2. Interelectrode Impedances Should Be Less Than 10,000 Ohms but More Than 100 Ohms

100

These are listed contraindications for HV

What is "recent intracranial hemorrhage, significant cardiopulmonary disease, sickle cell disease or trait, or patient inability or unwillingness to cooperate" (ACNS Guideline 1 3.8)

100

The minimum number of channels a standard EEG should have

What is 16 channels (ACNS Guidline 1)

100

This patient state is the priority to capture on EEG

What is sleep (ACNS Guideline 1, 3.9 and 5, 1.8)

200

This guideline explains listing electrode channels left over right and anterior over posterior.

What is ACNS Guideline 3: Proposal for Standard Montages to be Used in Clinical EEG

200

This header explains T7/8 and P7/8

What is ACNS Guideline 2: III. EXPLANATION OF THE MODIFICATION OF THE 10-20 NOMENCLATURE WITHIN THE MODIFIED COMBINATORIAL SYSTEM

200

Reduced sensitivity (10-15uV/mm) may be used in these cases

What is pediatric studies (ACNS guideline 5, 1.5)

200

This is the minimum distance a photic lamp should be placed from a patient's face.

What is 30cm.  ACNS Guideline 1, 3.8

200

This electrode is required for every EEG

What is a ground electrode (ACNS 1, paragraph 2.3)

300

This guideline explains the usage of T7/8 instead of T3/4, and P7/8 instead of T5/6

What is ACNS Guideline 2: Guidelines for Standard Electrode Position Nomenclature

300

This Table lists the recommended number of montages to use in each recording.

What is ACNS Guideline 3: TABLE 1: Number of Montages Recommended

300

These TWO types of patients should have a double distance EEG

What is a neonate with less than 30cm head circumference and a patient undergoing a brain death study (ACNS Guideline 5 and 6)

300

The minimum sampling rate required

What is 256 Hz (ACNS Guideline 4, Recording)

300

In this type of study, sensitivity must be increased to 2uV/mm for at least 30 minutes

What is an ECI recording (ACNS Guideline 6, paragraph 5)

400

This guideline lists the minimum sampling rate and display resolution

What is ACNS Guideline 4: Guidelines for Recording Clinical EEG on Digital Media

400

This header and paragraph number describes the standard filter settings for a routine EEG.

What is ACNS Guideline 1: 3 Recordings, Paragraph 3.4

400

This type of patient should have at least a 60 minute recording.

What is a neonate? (ACNS Guideline 5: Minimum Technical Standards for Pediatric Electroencephalography, paragraph 2.8)

400

The minimum common mode rejection

What is 90dB (ACNS Guideline 4, Recording)

400

This signal must be recorded at the start of each EEG

What is calibration (ACNS Guideline 1, 32. and 4, Notation of information during EEG recording)

500

This guideline calls for lowering the low frequency filter to 0.3-0.6Hz for certain recordings

What is ACNS Guideline 5: Minimum Technical Standards for Pediatric EEG

500

This header describes the requirement to include the number of electrodes and channels, whether or not video was used, and the patient's state on an EEG report

What is ACNS Guideline 7: Format for reporting: Technical Description

500

You must wait at least 24 hours before repeating this in pediatrics

What is an ECI recording (ACNS Guideline 6, 10. A Repeat EEG Should Be Performed When ECI Is in Doubt)

500

Number of montages to use during a live recording

What is 3 montages (ACNS Guideline 1, 3.7)

500

This guideline requires a "tap test"

What is an ECI study (ACNS Guideline 6, 3. The Integrity of the Entire Recording System Should Be Tested)