Procedural Requirements
Random Facts You Should Know
Equipment
Operation
Calculations
100
C-HP-1033.610
What is the procedure for Set-up and Operation of the AMS-4
100
Instrument Issue - bottom drawer
What is the response check source location?
100
2 Sealed Gas Proportional
What type of detector does the AMS-4 use?
100
a Satisfactory Calibration
What is required to be checked prior to use?
100
5.33e-9uCi/mL
What is product of 1e-8uCi/mL divided by 1.875 - Alternate DAC Calculation
200
32 to 122 degrees Fahrenheit
What is temperature range for operation of the AMS-4
200
0.3ft3/min
What is the minimum flow rate - flow rate value at which a minimum flow rate failure is determined.
200
Mario hits the blocks with this
What is his hand?
200
Listed on the Calibration Certificate as +/- 20% of stored efficiency
What is the Particulate Inline/Radial Efficiency Differential requirement?
200
It takes a sloth this long to travel one mile
What is a month?
300
Xe-133 or Kr-85 DAC
What is the Noble Gas setpoint limit based on?
300
There is enough iron in your body to make this
What is a 3 inch nail ?
300
"Wait for SAC / Wait for DAC"
What is the screen display when the AMS-4 is updating its history upon restart? (page 10 - tech manual)
300
2121
What is the default password?
300
Compares the count rate activity in the most recent 60 sec time interval to the previous one
What is the Fast Alarm Concentration Calculation?
400
Internal Exposure exceeds 10mRem (4 DAC-hours) during a calendar year
What is the basis for the DAC-hour setpoint?
400
Co-60 at 1e-8uCi/mL
What is the most restrictive DAC for the nuclides we expect and the basis for our Setpoint calculation
400
Audible Alarm with no Visual Alarm
What is a condition in which the AMS-4 has gone into Alarm but has now returned to normal levels
400
Coke would be this color if food coloring wasn't added
What is Green?
400
42.48LPM Air Flow Calibrator Reading
What is 1.5 CFM?
500
6.23e10
What is the factor used for calculating the CAM Setpoints - equals 2.2e6 x 2.832e4
500
This may cause increased levels of Kr-85 in the Fuel Building
What is a release of gap activity due possibly to failed fuel pins
500
A Brush and alcohol
What is the approved equipment/chemical for cleaning connectors?
500
Grab Air Sample, Change Patch, Monitor Conditions
What is the response to a CAM Alarm?
500
The instrument will count until the beta and background channel have at least 400 counts to provide a statistical accuracy of at least 10%
What is the Gamma Factor Count?
Continue
ESC
Reveal Correct Response
Spacebar
M
e
n