Pressure Systems
O2 Cylinder
Vaporizer
Cost and Amount Calculations
Hazards & Safety
100

What are the components of the high pressure system?

O2 source, regulator, pressure gauge

100

What is the most common O2 cylinder used in small animal?

E-cylinder

100

Main function of vapoizer?

Converts liquid agent to controlled vapor

100

Formula for liquid use per hr. 

Flow X % X 3 = mL 

100

What happens if vaporizer is tipped?

Contaminates bypass

200

What are the components in the intermediate pressure system?

Pipeline connections, flowmeter, O2 flush

200

What color is the O2 cylinder?

Green 

200

What type of vaporizer splits gas flow?

Variable bypass

200

Why reduce flow after induction?

Save agent and reduce cost (waste) 

200

What should you avoid putting on a tank valve and why?

Oil/grease=Explosion risk

300

What components make up the low-pressure system?


Vaporizer, fresh gas outlet, breathing circuit

300

How should the tanks be stored when not attached to the machine?

Upright and chained to a wall 
300

What is Isoflurane's vapor pressure?

240 mm/Hg

300

What is the constant for calculating liquid use?

300

How can a barotrauma occur?

Using O2 flush valve while connected

400

What does the flowmeter do?

Controls the flow rate of O2 
400

What is pressure (psi) and capacity (L) of the common small animal O2 tank?

1900 psi/ 660L

400
What does vapor pressure determine?

Determines the maximum deliverable concentration

400

If an E tank reads 560 psi and the flow is 2 L/min how much is left in the O2 tank?

560 / 1900 * 660=195L / 2 = 98 minutes 

400

What prevents the incorrect cylinder from being mounted to the machine?

Pin index system 
500

What does the regulator do?

Changes the O2 pressure from high to low 

500

What hangs the O2 tank to the machine?

Hanger yolk

500

What is the maximum deliverable concentration possible for isoflurane?

32%

500

What is the formula for calculating how much is left in the O2 tank?

Tanks current psi X 1900 psi X 660 L = answer / O2 flow rate

500

What are three things that could result in O2 supply issues?

Empty tank, tank disconnected, valve not open