Machine Parts
Cylinders
Airway
Monitoring
Hodgepodge
100

Oxygen flush valve type

What is a ball-spring valve

100

Liters in a full e-cylinder of Oxygen

What is 660L

100

What airway device is contraindicated for a "full stomach"

What is a LMA

100

This electronic device used by anesthesia providers monitors cognitive awareness during anesthesia 

What is Bispectral monitoring (BIS)

100

Responsible for the largest loss of heat during surgery

What is radiation- amt of body exposed to environment – loss of heat to the environment

Conduction - direct contact of body tissues or fluids with a colder material

Evaporative - latent heat of vaporization of water from open body cavities & the respiratory tract-sweating

Convection - loss of heat to air immediately surrounding the body

200

The first stage pressure regulator lowers the cylinder pressure to:

What is 45psi

200

How Nitrous oxide is stored

What is as a liquid

200

An advantage of LMA proseal

What is allows for higher positive pressure ventilation up to 30 cm h2o, has a gastric drainage tube access

200

This type of monitoring is definitive for recovery from non depolarizing neuromuscular blockade 

What is quantitative

200

This is indicated by the a wave in the CVP waveform

What is atrial contraction

300

Pressure reducing valve type

What is a diaphragm valve

300

PSI of a half full nitrous oxide tank

What is 745 psi

300

This special stylet helps with a difficult airway and can also be used as a tube exchanger

What is a Bougie or Eschmann

300

It is indicated by the downward stroke at the end of the CO2 waveform

What is inspiration

300

The I:E ratio that allows for a 2 second inspiration time at 10 breaths per minute

What is 1:2

60 sec/10 bpm= 6 sec/breath

6/3= 2sec 

400

Type of flowmeter tube

What is a Thorpe tube or a rotameter

400

Time remaining in oxygen tank at 950-1000psi with 3L running

What is 110 minutes

950-1000= 330 L

330/3=110

400

Type of ETT used during thyroid or parathyroid surgery

What is a Neural Integrity Monitor (NIM) tube

400

This is indicated by the slope on the pressure-volume loop

What is compliance

400

On the anesthesia machine, this ensures that FiO2 at the flowmeter can not be less than 25%

What is the proportioning system or N2O ratio controller

500

First internal component of the anesthesia machine that helps prevent delivery of a hypoxic mixture

What is the fail-safe valve

500

Entity responsible for the handling and transportation of gas cylinders

What is the Department of Transportation

500

Typical safe range for ETT cuff pressure

What is between 20-30 cm H2O

500

When in beach chair position the a-line transducer should be leveled and zeroed here

What is the circle of willis

500

The reason Desflurane vaporizer is heated to 39 degrees Celsius

What is the boiling point of Des (22.8 C) is close to room temperature - Tec 6 increases temp (39C) and pressure to 2 atm to keep Des a liquid. It uses electronic controls to precisely deliver des into the FGF