POLICIES AND PROCEDURES
WATER
DIALYSATE CONCENTRATES
SEQUESTERING
BAUDIT
100
THE SAMPLES TAKEN TWICE A YEAR FOR AAMI ANALYSIS
What is Tap Water and Last Water Outlet ( Return line acceptable )
100
ALL CARBON TANKS MUST BE FOLLOWED BY THIS SIZE CARTRIDGE FILTER ( OR SMALLER ) TO PROTECT RO MEMBRANES
What is 5 MICRON
100
DAVITA'S POLICY FOR MIXING RATIO OF DIALYSATE
What is 45X
100
POLICY FOR HANDLING MACHINE AFTER ALL TESTING AND VERIFICATION HAS BEEN PERFORMED BY BIOMED ( 3ITEMS )
What is DISINFECT MACHINE AND SEQUESTER MACHINE WITH A LOCK OUT DEVICE. LABEL DO NOT USE , SEQUESTERED
100
THE TWO NEW LINES ADDED TO THE BAUDIT AND WHAT THEY ARE ASKING FOR
What is LINE 6 : TOTAL CHLORINE TESTS MISSED AND LINE 7 : TOTAL CHLORINE TESTS DOCUMENTED CORRECTLY.
200
PER POLICY WE TAKE HOW LONG DOES IT TAKE TO VALIDATE WATER LOOPS
What is 4 CONSECUTIVE WEEKS
200
WHEN USING DI TANKS AS BACK UP PURIFICATION, THIS PRE TREATMENT COMPONENT WILL BE BYPASSED
What is THE WATER SOFTENER
200
THE MAJOR ELECROLYTE OF THE BODY'S BLOOD AND INTERSTITIAL FLUID
What is SODIUM
200
BIOMED WILL PERFORM AND DOCUMENT ON A BART TICKET THESE 5 MACHINES TESTS AND VERIFICATIONS
What is PRESSURE AND ALARM TESTS AND TEMPERATURE, CONDUCTIVITY AND pH READINGS
200
HOW MANY WORKING DAYS ARE BIOMEDS ALLOWED TO HAVE BAUDIT CLOSED AND SUBMITTED TO FA
What is 5
300
PER POLICY, HOW LONG DO WE HAVE TO REPEAT AN ELEVATED WATER SAMPLE ONCE RESULT IS RECEIVED
What is 48 HOURS
300
EBCT AND ITS REQUIREMENTS
What is EMPTY BED CONTACT TIME. 5 MINS EACH CARBON TANK
300
PER DAVITA POLICY, ACCEPTABLE RANGE FOR CONDUCTIVITY AND pH
What is 13 TO 15.5 and 6.9 to 7.6
300
THE THREE SAMPLES TO BE TAKEN FROM THE DIALYSIS MACHINE WHEN BEING SEQUESTERED..ALSO NAME THE COLOR TUBES
What is CULTURE AND ENDOTOXIN IN ORANGE TOP TUBE AND ELECTROLYTE IN RED TOP TUBE
300
NAME THE TWO SCOPES OF PERFORMANCE WE ARE MONITORING WITH THE BAUDIT
What is SAFETY AND COMPLIANCE
400
PER POLICY, HOW DO WE PREVENT INADVERTANT MIXUP OF ACIDS IN CENTRAL STORAGE SYSTEMS
What is USE OF AN ADAPTER
400
WHEN DI IS BEING USED AS PRIMARY PURIFICATION, RESISTIVITY MONITOR READINGS SHOULD BE RECORDED ON THE DI MONITORING LOG HOW OFTEN AND CONCURRENT WITH WHAT OTHER LOG
What is EVERY FOUR HOURS CONCURRENT WITH TOTAL CHLORINE CHECKS
400
WHEN USING A NON CENTRAL ACID DELIVERY SYSTEM, DIALYSATE CONDUCTIVITY AND pH TESTING IS REQUIRED THESE TWO TIMES
What is BEFORE EACH PATIENT TREATMENT AND WHEN CHANGING OR REPLACING THE ACID CONTAINER DURING THE COURSE OF A TREATMENT
400
DO NOT RETURN DIALYSIS MACHINE TO PATIENT USE WITHOUT PRIOR APPROVAL OF WHAT DAVITA DEPT. EXTRA CREDIT DAILY DOUBLE..NAME OUR LIASON FOR THIS DEPT
What is LEGAL RISK SERVICES.....CHRISTINE PERRY
400
WHEN REVIEWING LOGS FOR BAUDITS. WHAT IS THE POLICY FOR CORRECTING MISTAKEN ENTRIES ?
What is ONE LINE THROUGH INCORRECT ENTRY WITH BLACK PEN ALONG WITH INITIALS .
500
TO MINIMIZE BIOFILM FORMATION IN THE DISTRIBUTION SYSTEM, THESE ARE THE MINIMUM FLOW VELOCITIES PERMITTED AT THE DISTAL PORTION OF DIRECT AND INDIRECT FEED SYSTEMS ( 2 ANSWERS)
What is 3 FT/SEC FOR INDIRECT AND 1.5 FT/SEC FOR DIRECT UNDER PEAK DEMAND
500
THE RO WILL BE EQUIPPED WITH MONITORS THAT PERMIT THESE TWO TYPES OF MEASUREMENTS OF THE PRODUCT WATER ( ALSO GIVE THEIR UNIT OF MEASUREMENT )
What is QUALITATIVE ( CONDUCTIVTY OR TDS ) AND QUANTATIVE ( GAL/ PER MINUTE )
500
5 ELECTRLOYTES IN DIALYSATE CONCENTRATE
What is POTASSIUM, MAGNESIUM,GLUCOSE,CHLORIDE, AND CLACIUM
500
NAME TWO OTHER SITUATIONS WHEN DAVITA ALSO RECOMMENDS SEQUESTERING DIALYSIS EQUIPMENT
What is PYROGENIC REACTION AND AIR EMBOLISM
500
ONE OF THE QUESTIONS ON THE BAUDIT IS " ARE pH AND CONDUCTIVTY TESTING BEING PERFORMED. THESE ARE THE PARAMETERS FOR PHOENIX METER VS STANDARD SOLUTION AND PHOENIX METER VS DIALYSIS MACHINES. DAILY DOUBLE : WHAT IS THE TOLERANCE FOR THE MYRON L D1 OR D2 METER ?
What is =/- .1 AND =/- .4 WHAT IS 0