Equipment
column 2
Limits of error
Column 4
Column 5
100

Collaboration among staff for equipment selection.

Team Decision-Making

100

The same factors should produce consistent out put. This is called?

Exposure reproducibility

100

what is the limit of error for mR/mAs and linearity?

+- 10%

100

Beam alignment should be with _________ cm 

1

100

Create 2D images for focal spot assessment.

Pinhole Cameras

200

System for monitoring equipment performance and standards.

Quality Management (QM)

200

Test that can be done for Focal spot size (FSS)

Line pair resolution

star test patterns

pinhole camera

slit camera

200

What is the limits of error for reproducibility?

+-5%

200

Activities ensuring consistent high-quality radiology services.

Quality Assurance (QA)

200

Produces 1D images for small focal spot sizes.

Slit Camera

300

Parameters needed for effective imaging equipment selection.

Imaging Requirements

300

What % must AEC sensitivity be within?

+-10%

300

Limit of error for kVp accuracy?

+-5%

300

Monitoring equipment to maintain quality standards.

Quality Control (QC)

300

Data collection process for analyzing poor- quality images. 

Rejected image analysis

400

Detailed requirements based on team input for equipment.

Equipment Specifications

400

The exposure rate in fluoro may not exceed:

5R/min

400

Limit of error for tube angle accuracy?

+-1%

400

Mandated standards like HIPAA for patient confidentiality is considered?

Government Regulation

400

Source to skin (SSD) to mobile device

12"

500

Comparing bids based on price, function, and service.

Selection of Equipment

500
What is a normal repeat rate ?

3-10%

500

Limit of error for light field and central cross hairs?

light field +-2% of SID

cross hairs +-1%

500

Detects discrepancies between exposures at different mAs.

Artifact Identification

500

Source to skin (SSD) for stationary device  

15"