What does it mean if the corresponding light for a consumable bottle (ex. Procell, Cleancell) isn't lit up green?
This consumable is in-use by the analyzer.
When was Sample 00032 pipetted?
06/10/2026 11:43:29
What is included in daily maintenance?
Green wash rack and Cleaning the probes on the 801 side.
How can I see my recent calibration results?
Routine > Recent Calibrations
What are the rack ranges for Standard tube - Whole Blood samples?
racks 21-28
Can a medical lab scientist change the measuring cells on the e801 unit?
No, only Roche Engineers can change the measuring cells.
Print a sample report
Show on screen
Which solutions are used for the daily and weekly wash rack?
Bonus: What positions are they in?
Activator and Sysclean
Daily: positions 2(sysclean) and 3(activator)
Weekly: 1&2 (sysclean) and 3 (activator)
No
How do you properly put the analyzer in standby? shutdown?
Deselect rack reception.
Power button.
Which consumable bottle on the e801 has a RFID tag?
Procell
A result didn't cross over into Epic, how do I resend it from the analyzer?
Show on screen
What do you clean the reagent probes with on the e801?
Alcohol prep pad and water
Does Immunoassay QC have to be assigned to a rack?
No they have barcodes.
Are the Cobas link and Host connected?
Yes
How long do you hold the reagent loader button if you have MORE than 5 reagents to load?
short press or less than 2 seconds.
Make a filter for Processing Quality Control samples
Show this on screen
How often do we change the photometer lamp?
750h or every 6 months
When was the last time CHOL was calibrated?
6/10/26
How do you create a CMP Profile?
Menu>System>Key Assignment>Configure test profile
What is the S2 probe on the c503 used for?
Whole Blood samples for Hemoglobin A1C
Why does sample 00078 have a paper symbol under the status indicator?
Bonus: What is the technical limit?
> than the technical limit, needs a dilution
1.40-112
How is Ecotergent made?
1:50 dilution of Eco-D and DI water
When is a lot calibration generated? Reagent pack calibration?
Lot calibration is when a reagent has been loaded for <24 hours. Reagent pack cal >24 hours.
How many hours are left on our photometer lamp?
show on screen