Analyzer
Results
Maintenance
QC/Calibration
Software
100

What does it mean if the corresponding light for a consumable bottle (ex. Procell, Cleancell) isn't lit up green? 

This consumable is in-use by the analyzer. 

100

When was Sample 00032 pipetted? 

06/10/2026 11:43:29

100

What is included in daily maintenance? 

Green wash rack and Cleaning the probes on the 801 side.

100

How can I see my recent calibration results? 

Routine > Recent Calibrations

100

What are the rack ranges for Standard tube - Whole Blood samples?

racks 21-28

200

Can a medical lab scientist change the measuring cells on the e801 unit? 

No, only Roche Engineers can change the measuring cells. 

200

Print a sample report 

Show on screen 

200

Which solutions are used for the daily and weekly wash rack?

Bonus: What positions are they in?

Activator and Sysclean 

Daily: positions 2(sysclean) and 3(activator)

Weekly: 1&2 (sysclean) and 3 (activator)

200
Can yellow and green calibrators go on the same rack?

No 

200

How do you properly put the analyzer in standby? shutdown? 

Deselect rack reception.

Power button.

300

Which consumable bottle on the e801 has a RFID tag?

Procell

300

A result didn't cross over into Epic, how do I resend it from the analyzer? 

Show on screen 

300

What do you clean the reagent probes with on the e801?

Alcohol prep pad and water 

300

Does Immunoassay QC have to be assigned to a rack?

No they have barcodes.

300

Are the Cobas link and Host connected?

Yes 

400

How long do you hold the reagent loader button if you have MORE than 5 reagents to load? 

short press or less than 2 seconds. 

400

Make a filter for Processing Quality Control samples

Show this on screen 

400

How often do we change the photometer lamp?

750h or every 6 months

400

When was the last time CHOL was calibrated?

6/10/26

400

How do you create a CMP Profile?

Menu>System>Key Assignment>Configure test profile

500

What is the S2 probe on the c503 used for? 

Whole Blood samples for Hemoglobin A1C

500

Why does sample 00078 have a paper symbol under the status indicator?

Bonus: What is the technical limit?

> than the technical limit, needs a dilution 

1.40-112

500

How is Ecotergent made? 

1:50 dilution of Eco-D and DI water 

500

When is a lot calibration generated? Reagent pack calibration? 

Lot calibration is when a reagent has been loaded for <24 hours. Reagent pack cal >24 hours.

500

How many hours are left on our photometer lamp? 

show on screen 

M
e
n