Saline Prime
Air Alarm
Transducer Alarm
Conductivity Alarm
Lagniappe
100

How would you trigger a port error?

By opening the substitution fluid port or rinse port door during test.

100

What alarm will you see if there is a low venous chamber and how do you correct the alarm?

Air alarm- by raising the level on the Blood System Menu

100

What does the transducer protector do?

Protects the machine from blood contamination while transmitting pressure.

100

What does conductivity measure?

The concentration of electrolytes in the dialysate.

100

Which alarm protects the patient when tubing is not secured into the optical detector?

Air alarm

200

If a saline setup is used because of an Online port error, what must be completed before the patient's next treatment?

Chemical Heat Disinfection

200

What are somethings you should do when you have an alarm?

Assess the patient, silence the alarm, and you the assist mode.

 
NEVER ignore the alarm!!!

200

What usually causes a transducer alarm?

A wet transducer

Clot formation

Line occlusions

200

When is it a good time to change the bi-bag in an ongoing treatment?

When there is 30 minutes or more left within the treatment.

200

Which alarm verifies the dialysate is properly mixed?

Conductivity alarm

300

If saline is not an option on the Blood System Menu, what do you do?

Trigger an Online port error

300

What are some possible causes of an air alarm?

Loose connection

Venous line not secured into the optical or air detector

Air introduced during medication administration


300

When you have a transducer alarm what should you inspect for?

A wet transducer or clots

300

What are some causes of a conductivity alarm?

Incorrect or empty acid 

Incorrect or empty bicarbonate

Occluded or kink connections


300

Which alarm is associated with a wet pressure monitoring system?

Transducer alarm

400

What are two things you should inspect for during saline prime?

Air

Leaks

Secure connections

400

What can happen to the patient if air gets into their blood system.

Air Embolism

400

What are some things you must do in order to replace the transducer?

Don full PPE

Remove the fluid in the venous pressure chamber

Use hard keys to stop the blood pump, open EBM doors

Clamp off the venous pressure monitor line with blue hemostat clamp.

400

What patient complication can occur if conductivity is incorrect?

Hemolysis

Electrolyte imbalance

Cardiac arrhythmias

400

What is one common safety principle for all machine alarms?

Never bypass or ignore an alarm

500

Tell me how to turn the T1 test on if it does not go into test automatically?

Dialysate Menu

Turn the Flow I/O on

500

Should the blood pump be restarted before removing the air?

No 

500

What are some steps that must be ensured when changing the transducer?

Keep the old venous pressure line in place but secure clamp out of way to prevent blocking the blood pump

Start blood pump and ensure the level of venous chamber is reset.

500

How can a conductivity alarm be fix?

Ensure that the bibag is not empty.

Verify that the acid is correct.

Verify no loose connection or kinks.

500

What should be verified for the patient before starting a treatment?

Verify that the correct acid and bicarb is used for the patient