Hemodynamics
Gen. Doppler
Color Doppler/Power Doppler
Optimization
Continuous Wave
100

Name 1 of the 5 Effects of a Stenosis 

1. Change in Flow Direction 

2. Increased Velocity as The Vessel Narrows

3. Turbulence Downstream from The Stenosis

4. Pressure Gradient Across The Stenosis

5. Loss of Pulsatility 

100

Most Common Error Associated with Doppler Ultrasound 

Aliasing 

100

Two most common color maps

1. Variance mode

2. Velocity mode

100

You are imaging a kidney. Your color box is set on the kidney, but no color is appearing. Your color gain is set to 5.0 The image is sector shaped, so we dont need to consider that the sound beam and flow are perpendicular. How do you fix it? 

Turn your color gain up! 

100

Continuous wave doppler requires how many crystals for imaging

2

200

What describes the relationship between velocity and pressure in a moving fluid? 

Bernoulli’s Principle

200

What eliminates low frequency Doppler shifts near the baseline of a spectral display but has no affect on high frequencies?

Wall Filters/ High pass filters

200

True or False: Color Doppler measures peak velocity.

False. Color Doppler measures mean velocity and Spectral Doppler measures peak velocity. 

200

You are imaging an adjacent pair of vessels. One is an artery and one is a vein. You notice splashes of color outside of the artery posterior to the wall.  How would you fix this? 

Increase the Wall Filter. It removed color from slowly moving reflectors such as blood cells and vibrating tissues. 

200

T or F: A dedicated continuous wave transducer creates anatomic images.

False, the transducer only contains two elements, one transmits continuously while the other receives.

300

Unitless number that determines whether flow is laminar or turbulent

Reynolds Number

300

Fast Fourier Transform Distinguishes What Two Things? 

Laminar from Turbulent Flow

300

Name one of the two advantages of  Larger Color Doppler Packets.

1) More Accurate Velocity Measurement

2) Increased Sensitivity to Low Flow 

300

To be able to select the most optimal location where velocity is measured is 

Range resolution/ Range specificity 

300

What is a major disadvantage of CW Doppler 

Range Ambiguity 

400

Three types of energy loss associated with hemodynamics 

1. Frictional Loss

2. Viscous Loss

3. Inertial Loss

400

What is the cosine of a 120 degree Angle? 

-0.5

400

What is the digital technique used to analyze color flow Doppler because of the enormous amount of data that is processed?

Autocorrelation 

400

T or F: if the colors that touch each other in the map go through the middle, aliasing is present. If the colors that touch each other go around the outside of the map, flow reversal is present. 

FALSE! 

400

The greatest advantage of Continuous Wave Doppler is its ability to what? 

Accurately measure very high velocities. 

500

Venous Flow From What 3 Things Increases During Inspiration

The Head 

The Arms 

The Vena Cava 

500

What does the X-Axis and Y-Axis represent on a Doppler Spectrum? 

X-Axis = Time

Y-Axis = Doppler Shift/Velocity

500

Name 1 advantage and 1 disadvantage of using Power Doppler. 

Advantages: increased sensitivity to low flow or velocity, unaffected by Doppler angles (unless exactly 90 degrees), no aliasing. 

Disadvantages: no measurement of velocity, lower frame rates, susceptible to motion of the patient, transducer, or soft tissues (causing flash artifact). 

500

You are optimizing a color Doppler image. You notice that red is showing flow away from the transducer and blue is showing flow toward the transducer, but you expect the opposite. The color box is also over the whole image, and there is excessive color noise throughout the image. What three adjustments should you make to optimize the image?

1. Invert the color map 

2. Decrease the size of the color box. (Improve Frame Rate)

3. Decrease the color gain. (Reduce Color Noise)

500

Dedicated Continuous Wave Transducers do not use backing material, which results in the following 4 things: 

1) Undampened Transmitted Signal 2) Narrow Bandwidth 3) ? 4) ? 

Name one of the other two. 


High Quality Factor or Higher Sensitivity 