Eclipse General
Halcyon 2.0
Halcyon 2.0 Planning
Halcyon 2.0 Planning Part 2
Halcyon Planning Part 3
100

What are the two main new features for Eclipse 15.6?

1. Treatment Planning Support for Halcyon v2.0

2. Fast MCO Hybrid VMAT


100

What is the old name of Halcyon Planning System?

RDS (Ring Delivery System)

100

What are the 6 new features in Halcyon v2.0 planning?

  1. Long Fields

  2. Dynamic Beam Flattening

  3. kVCBCT imaging support

  4. Delta Couch Shift one step setup

  5. AAA enhancements (Tongue and Groove)

  6. Machine change capability







100
  1. What is the maximum size of length that can be treated on Halcyon v2.0?

Ans: 36cm

100

True or False

  1. Plan templates and Clinical Protocols do not support Multi-isocenter plans. This will be supported in the next version of Eclipse.

False

200

True or False

Eclipse 15.6 gives you the ability to edit Structure codes for all structures in a structure set at once. You do this in the properties>summary>structure code of a structure set.

True

200

What energy options are available for Halcyon?

6X FFF

one pre-configured model only

200

True or False

You can edit the size of the Halcyon kV CBCT image in the planning system

False

200

True or False

The maximum separation of the treatment isocenters is 4cm for Halcyon Long fields in the longitudinal direction

False

The maximum separation between treatment isocenters is 8cm

200

True or False

  1. It is now possible to create 3D plans for the Halcyon Machine v2.0 

True

By using Dynamically Flattened Beams

300

Where are the buttons to quickly change the lay out from 3D to BEV located?

In External Beam Planning> Model View

300

What 6 treatment delivery techniques does Halcyon v2.0 support?

1. VMAT

2. IMRT

3. FIF

4. Irregular Surface Compensator

5. Dynamically Flattened Beams

6. Extended Fields

300
  1. What is the maximum number of treatment isocenters allowed for a long treatment field on Halcyon v2.0?

Two

300

Eclipse now has the ability to limit leaf movement in the X direction for Halcyon VMAT fields.

  1. Where do I set these limits?


  1. Plan Information drawer of the PO

300

Can Halcyon plans be created/edited in Plan Parameters?

No
All plans must have dose calculated for approval.
Only non dosimetric edits can be made such as
Field, ID, Name, Tolerance, SSD

400

What change was implemented in the Arc Geometry Tool?

Arc Geometry Tool now supports Halcyon Multi-isocenter plans

400

How is the field size defined in Halcyon and what is it?

Defined by the MLCs (No jaws)
28 x 28 fixed Field Size

 


400
  1. The New SX2 MLC effectively gives you a leaf resolution of:

0.5cm

400

True or False

  1. If delta shifts have been calculated in Eclipse, they can be applied on the Halcyon v2.0 machine in one step setup selecting the Load button on the machine. Ensure that the shifts are not also being applied manually as for Halcyon v1.0

True

400

Where is time entered in the Eclipse and RTM Workspaces?

Nowhere: it is computed at the treat unit.
It is not entered as part of the field parameters at all.

500

What does the Fast MCO Hybrid VMAT feature do and how does it work?

1. Makes plan generation with MCO for VMAT planning faster

2. Converts the VMAT plan into surrogate IMRT plans that are used for plan generation in MCO trade off exploration

500

A. Describe three things about the Halcyon MLC configuration.

B. What is the proximal layer closer to?

1. 2 sets of multilayered 1cm leaves
offset from each other by 0.5 cm

2. MLCs rotate (Primary Collimator does not)
3. MLC leaves travel the full extent of the field (no carriage split)

B. Closer to the source

500

What MLC layer is used to flatten the Halcyon v2.0 beam?

Proximal layer

500

Tell me three things about imaging fields for Halcyon

1. Automatically created when adding new plans
2. MV image dose is included in optimization/ calculation
3. MV dose accumulation is automatically accounted for with imaging


500
  1. With Eclipse 15.6 and Halcyon v2.0, you can now convert the following:

Convert Halcyon to a non-Halcyon machine with 120MLC or HDMLC or vice versa

M
e