Inhalational Agents + Vaporizers
Electrical + Fire Safety
Neuromuscular Monitoring
Anesthesia Machine
Miscellaneous
100

Property of a substance/material that determines how fast it transmits heat (pertinent to the material of anesthetic vaporizers to reduce fluctuations in temperature)

thermal conductivity

100

FiO2 limit during procedures with a high airway fire risk 

FiO2 should be <30%

100

Dibucaine number associated with homozygous atypical pseudocholinesterase and the effects of succinylcholine can be greatly prolonged by several hours

20-30

100

Most common type/classification of bellows because if a circuit disconnect occurs, the bellows will NOT be able to rise during expiration

ascending bellows

100

Two blood products that should NOT be administered through a fluid warming device

platelets + cryoprecipitate

200

In clinically equivalent doses, list the negative climate impacts of the inhalational agents (from most negative to least negative)

desflurane > N2O > isoflurane > sevoflurane

200

Components of the Fire Triad and one example of each 

fuel - ETT, surgical drapes, alcohol preps

oxidizer - O2, N2O

ignition- electrocautery, lasers

200

Percentage of nicotinic acetylcholine receptors (nAChR) that may still be blocked by a nondepolarizing NMB with a train-of-four count (TOFC) of 4

70-75%

200

Essential safety feature of the anesthesia machine that prevents contamination of the OR with waste anesthetic gases

scavenging system

200

Name one valuable thing that you learned from this class and that you will remember forever 

:)

300

Two properties of desflurane that cause it to require a special type of vaporizer 

high vapor pressure + low boiling point

300

In the event of the line isolation monitor (LIM) alarming that faulty equipment is present, what action should be performed?

UNPLUG the last line that was plugged in***; continue to unplug lines until the alarm stop

300

Muscle monitored with a peripheral nerve stimulator that has the slowest recovery time from paralysis by nondepolarizing NMBs

adductor pollicis m. 

300

Essential safety feature of the anesthesia machine that ensures a minimum concentration of oxygen is delivered to the pt, it links the flow of two medical gases 

oxygen/nitrous oxide ratio controller

300

Etomidate induction dose

0.1-0.4 mg/kg

400

3 common characteristics of vaporizers (NOT for electronic vaporizers)

variable-bypss, temperature-compensated, agent-specific, flow over

400

Maximum current leakage allowed in OR equipment

10 µA

400

neostigmine + glycopyrrolate doses for reversal of neuromuscular blockade by nondepolarizing NMBs

neostigmine 40-80 mcg/kg

glycopyrrolate 0.2mg for every 1 mg of neostigmine

400

Two characteristics of a circle breathing system that make it more desirable than a Mapleson circuit

better control of anesthetic depth, ability to scavenge, conservation of heat + humidity, rebreathing of exhaled anesthetic gases (less waste)

400

Three indications for the use of a total intravenous anesthetic (TIVA)

Malignant hyperthermia susceptibility, neuromonitoring (use of motor evoked potentials (MEPs)), ENT procedures without a secured airway, intracranial procedures, hx of significant/severe PONV

500

Pressure + temperature at which an electronic vaporizer pressurizes and warms desflurane to

The electronic vaporizer pressurizes desflurane to ~1500 mmHg and warms it to ~40 ºC

500

Gauge reading on the line isolation monitor (LIM) that can indicate faulty equipment is present

> 5 mA

500

Two keys differences between myasthenia gravis (MG) and Lambert-Eaton myasthenic syndrome (LEMS)

MG: autoimmune rxn againts post-synaptic AChR, muscle function worsens with repeated activity, symptoms start proximal and progress peripherally, resistance to succinylcholine + sensitivity to nondepolarizing NMBs

LEMS: autoimmune rxn against pre-synaptic calcium channels, muscle function improves with repeated activity, symptoms start distal and progress proximally, sensitivity to both succinylcholine + nondepolarizing NMBs

500

Components of a circle breathing system

•CO2 absorber

•Fresh gas inlet

•Inspiratory + expiratory unidirectional valves

•Inspiratory + expiratory breathing tubes

•Y-connector

•APL valve

•Reservoir bag

500

Epinephrine infusion rate range

0.01-0.1 mcg/kg/min

M
e
n