What temperature do you store pH/Gas sensor cards?
2 - 8C
How long does it take for the Chemistry sensor card to hydrate?
2 hours
What temperature is the Chemistry Auto QC stored?
-20C
What parameters are dependent on temperature?
pH, pO2, pCO2, O2%, CO2%, HCO3
Where do you order new raw materials for the Flex2? (E.g. probes, pump harness, etc)
2023 Material Tracker under Raw Materials on Teams
How many parts are there in the pH/Gas module as a whole?
3. Sensor card, cartridge, AutoQC
How do you change the chemistry calibrator cartidge? Detail all steps.
Activate by removing the clips, pushing the internal cup through the crease that was created by the clip and close the flap. Shake vigorously to mix for NLT 2 minutes. Position new pack within its port and push it back, ensuring that the back rubber filaments are pierced by the needle filaments.
What should you put on the rubber septum of the pH/Gas Auto QC when replacing it?
Drop of water on both septums.
What is the default temperature of the Flex2?
37C
The osmo module turns yellow and there are plenty of tubes. What do you do?
Make sure new tubes are in slots 1 - 10. Calibrate the Osmo with two standards
PO2 is not passing QCs and the pH/Gas sensor card is nearing expiration. What do you do?
Change the sensor card.
When disposing of the chemistry cartridges, what do you write on the hazardous waste sticker?
Dilute Solution of alkaline and alkali metal salts, buffers and preservatives
How long should you wait to replace the pH/Gas cartridge after you take it out of the fridge?
24 hours
Upper right hand corner. i in a circle -> Video help
How do you calibrate the Osm20?
First aspirate Osm50 standard and then aspirate Osm850 standards.
Your pH/Gas Auto QC cartridge is faulty and you still need to use the Flex. What do you do?
Run external QCs Level 1, 2, and 3.
-20C
"No Sample Acquired" when running pH/Gas Auto QCs. What do you do?
Take out the pH/Gas Auto QC, push a needle through both rubber septums in the back of the cartridge. Reinsert with firm pressure.
What is the difference between Osm20 and Osm48?
Amount and shape of tubes, no wiper ring for Osm48
What are the different parts you have to change when replacing the ESM cartridge?
ESM cartridge (outside), rubber septum (inside), little plastic tubing (inside)
You discovered a failed sensor on your pH/Gas sensor card. What should you do?
Claim the warranty, contact Nova with the warranty number and the serial number of the analyzer.
What is the PD number of the chemistry cartridge?
Osmo is locked out. You run Auto QCs and no result appears for Osmo. What do you do?
Turn off QC lockout and rerun onboard QCs. Once everything passes, turn QC lockout back on.
What is the sparging O2% default?
20.9
If you need to run module depro, do you need to replace the sensor cards after or can you place the cards back in?
You do not need to replace the sensor cards with new cards. You can place the old cards back in if samples still remain.