500

Suggested external review on first implementation of a program, such as by this TLD/OSLD review service. Coincidentally, something people tell me all the time...

IROC

500

According to TG 178, the optimal chamber orientation for Gamma Knife

Along the Z-axis.

Placing the chamber along other axes could cause tangent effects, producing a lower measurement

500

Collimator size that should be used for monthly outputs, beam shape, and beam symmetry measurements

60 mm (the largest available collimator)

500

DQA frequency of Cyber Knife

Due to complexity of testing non-isocentricity:

-1st several patients for new tracking modality

-monthly

1000

Name given to the dose error caused by the opening and closing of the source window for a gamma knife

shutter effect

D(t) = \dot(D)(t'-\tau)

1000

spherical, calibration phantom from elekta, made of Acrylonitrile Butadiene Styrene

ABS phantom

1000

How do you test the tracking ability of the cyber knife?

use ball sphere and anthropomorphic phantom, move phantom set distance w/in 10 mm

1000

Frequency with which a physicist should observe a Cyber Knife's Synchrony treatment, whether it be on a patient or phantom

Monthly

Synchrony is the motion tracking feature. Puts a lot of stress on robot joints, especially when interrupted and the brakes have to engage often. Check for unusual noises or vibrations

1500

meaning of msr found in the following equation: 

D_{W,Q_{msr}}^{f_{msr}}=M_{Q_{msr}}^{f_{msr}}\times N_{D,W,Q_0}\times K_{Q,Q_0}\times K_{Q_{msr},Q}^{f_{msr},f_{ref}}

machine-specific reference

1500

test to confirm radiation focal point (RFP) and unit center point (UCP)

place film in film holder, attach to patient position system. Measure x/z and y/z (2 films for each coll.)

1500

A test analogous to a Winston-Lutz to be performed daily. Has a phantom of a similar name. 

Starts with CT of the phantom, QA plan including an AP and lateral beam, and measurement of the relative position of the concentric circles produced.

AQA test (automated QA)

1500

gold standard detector for TPR, Off-center ratio, and output factor annual measurements

Diodes

-ion chambers, even microchambers, not recommended below 20 mm collimator

2000

Frequency Cobalt-60 sources are replaced

every 5-10 years

2000

What is this measuring:

{{dD}/dt(4)}/{{dD}/dt(16)}

Output factor for a 4 mm collimator

2000

Static isocentric E2E test for a well calibrated cyber knife, should result in what level of accuracy

0.3-0.7 mm. 

This test involves imaging a ball cube phantom, creating an isocentric plan with the sphere covered by the 70% isodose line and comparing the centroid between plan and measurement

2000
DQA (delivery QA) acceptance criteria

90% acceptance down to 50% IDL.

- 2%/2mm

-3%/3mm for Synchrony

1000000

Name of the recommended formalism for Gamma Knife calibration

ALF (Alfonso)