Inlets
Detectors
Potpourri
Site Prep/I & F
Potpourri II
100
This valve controls total flow in the SSL
What is PV1 or Proportional Valve 1
100
This FID base requires the use of a "long" jet.
What is Capillary Adaptable Base.
100
You can find all of the gas, space, and power requirements here.
What is the Site Prep document on the GC page on the Support Portal
100
This ships with all GC systems and is what you use to guide the customer through familiarization.
What is the eFam disk for GC Hardware.
100
You could use this mode on the SSL if you were exceeding the liner volume and could not change the liner type.
What is Pulsed Splitless or Pulsed Split mode.
200
If you configure a 320u column to a PP Inlet it MUST be in ______ ______ mode and then _____ _____ mode for a Packed Column.
What is Pressure Control and Flow Control mode.
200
This makes up the difference in flows that comes with the use of capillary columns in both the FID and TCD.
What is Makeup Gas.
200
This inlet can have a septumless head design.
What is the PTV.
200
This should be completed and signed by you and the customer at the time of install.
What is the Installation Checklist.
200
Using this makeup gas for the FID may give you higher sensitivity.
What is Nitrogen.
300
If this valve in the Split Splitless Inlet fails, it will cause the Inlet to go to Line Pressure
What is PV2 or Proportional Valve 2
300
If you TCD is "ringing" then you need to do...
What is change the data collection rate to 5Hz
300
The 4 physical measurements you make during an OQ service and the tools associated with them.
What is Continuity (Voltmeter), Pressure (Manometer), Flows (Flowmeter), Temperature (Temp Meter/Probes).
300
This must be measured and within spec prior to installing the GC
What is Line Power.
300
You could do this un-invasive test if you had trouble igniting the flame.
What is the Jet Diagnostic or Jet Restriction test.
400
When making a Large Volume Injection you are operating in this mode.
What is Solvent Vent Mode
400
The FID collector is at "charged" at what Voltage and why
What is ~-220V to attract the + charged ions that are formed as a result of being burned in the FID flame.
400
Has the sample capacity for 16 vials.
What is the High Density Turret on the 7693 ALS.
400
A GC with a Gas Sampling Valve requires the use of this.
What is a separate 50psig Air source.
400
The MMI has a _______ for a heater and is ______ watts.
What is THERMOCOUPLE and 150.
500
This valve that is NOT part of the EPC module (pressure/flow control) is mounted in that same space.
What is the CO2 cryo valve.
500
The default lit offset for the FID
What is 2.0 pA
500
You change this part when you are going to be making Large Volume Injections.
What is the Syringe Carriage.
500
The minimum supply pressure for inlets on the GC.
What is 20psig higher than the method.
500
The SSL inlet requires a column installation of __-__ mm and the MMI requires a column installation of __-__ mm but the PP inlet requires only __-__mm, why?
What is 4-6, 10-12, and only 1-2 because you don't want any un-swept dead volume since everything needs to go onto the column in the PP.