Computed Radiography
Image Intensified Fluoroscopy
Digital Fluoroscopy
Digital Image Processing
Quality Control
100

The photostimulable phosphor used in computed radiograpic imaging plates.

What is europium activated barium fluorohalide

100

The input phosphor is made of

What is cesium iodide
100

The scintillator in an indirect capture CCD imaging system is 

What is cesium iodide

100

A graph of the number of pixels of each gray value in the digital radiographic image

What is a histogram

100

Leakage radiation must be limited to less than

What is 1 mGy/hour at 1 meter from the source

200

The random, disorganized crystals in an image phosphor layer.  

What is turbid

200

The component of an image intensifier that converts light to electrons is composed of

What is cesium and antimony (in the photocathode)

200

The photodetector in an indirect capture TFT imaging system 

What is amorphous silicon

200

The preprocessing function that adjusts a radiographic image for over or under exposure

What is automatic rescaling

200

The bucky slot cover and protective curtain must contain at least

What is 0.25 mm Pb equivalence 

300

Where the electrons from the ionized phosphor crystals are "trapped" in the imaging plate.

What is the conduction band

300

The brightness gained due to the crowding of light photons onto the output phosphor 

What is minification gain

300

The photoconductor in a direct capture TFT imaging system

What is amorphous selenium

300

The S value that the exposure indicator is based upon.

What is S(ave)

300

The image intensifier housing must contain at least

What is 2.0 mm Pb equivalence

400

In the CR reader, light photons are released from the imaging plate with a ________ and collected by a PMT or CCD.   

What is a red laser

400

The elements in the image intensifier that converts electrons into light

What is silver activated zin cadmium sulfide (output phosphor)

400

What an indirect capture DR system has that a direct capture DR system does not

What is a scintillator

400

This process controls the contrast scale in a digital image

What is LUTs (look up tables)

400

The source to skin distance for a portable fluoroscopy unit must be at least

What is 12 inches (30 cm) 

500

The rate that data is acquired from the imaging plate is called the ____.

What is sampling frequency

500
Fluoroscopy tubes operate under this mA

What is 2-5

500
CMOS does not require this device that is required in a CCD DR imaging system.

What is an ADC

500

A "spike" on the right on a visualized histogram represents

What is "raw radiation"

500

Fluoroscopy exposure at tabletop for a standard ABS image intensified system is limited to 

What is 100 mGy/min