X-Ray Equip Review
Radiographic Visual/Enviro Inspections
Radiographic Test Tools
Safety Code 35
Radiographic Artifacts & Repeat Analysis
100

Full Wave Rectifiers generate ________ pulses per second

What is 120 Pulses per second

100

The control console inspection includes a visual of _____________

What are panel lights, switches and tube heat sensors

100

Beam-Bucky alignment can be tested utilizing what two testing methods?

1. Collimator Test Tool 

2. Washer/Coin Method

100

Positive Beam Limitation Systems must be within 

3% SID for either length or width

Combined length/width error </- 4% SID

100

Purpose of performing repeat analysis is

- to identify common causes

- improve image quality, workflow and staff performance 

200

Half-Wave Rectified Generators are still used in?

What are dental offices?

200

The Overhead Tube Crane Assembly visual inspection includes_________________________

What is high voltage cables, lock functions, SID/angle accuracy, collimator light

200

Digital Displays are evaluated utilizing what? 

SMPTE Test Patterns/ AAPM (TG18)

200

X/Y Collimator Accuracy must be within:

+/- 2% of SID

200

List three common causes of radiographic artifacts.

1. Motion

2. Operator Error (grid lines, tube angle)

3. Failure to remove clothing, jewelry etc.

300

The filament transformer is a _________ transformer

What is a step down transformer?

300

Mechanical Visual Inspections for electrical integrity includes___________________________

High-tension Cable insulation (discoloration, loose retaining rings)

300

What key item is utilized in determining Radiation Output reproducibility? 

Dosimeter

300

At 110 kVp, what must the HVL be?

3.9 mm of aluminum 

300

In DR systems, ghosting is caused by _______________

Residual charge within the TFTs

400

Three-phase twelve-pulse generators produce _________ voltage ripple. 

What is 3.5%

400

Visual inspections of the patient table include _________________________

Table surface condition/cleanliness, function of angulation/power switches, bucky tray and tabletop material is less than </- mm Al equivalent

400

SNR is a measure of what radiographic concept?

Image Noise

400

AEC testing should be completed ______________

At installation and annually 

400

Scratches and dirt show up how, on radiographic images utilizing CR plates?

Increased density (white) creases and specks
500

Power Rating determines_________________

What is the Xray systems maximum power?

500

Miscellaneous equipment that requires visual inspection includes ________________________

Positioning sponges, patient aids, step stools, IV poles 

500

What is used to determine the HVL and how much is generally required

6-8 mm aluminum plates

500

Radiation output variability must not exceed _____

5%

500

Failure in radiation output reproducibility would result in ______________________ 

Inconsistent image quality, increased repeat exposures, failure to meet SC 35

M
e
n