Back To Basics
True or False
Scope Reprocessing
Yikes, I'm tachycardic
Duodenoscopes
100

How many times should the yellow sponge be squeezed once placed in the eco bowl?

At least 3 times 

100

Simethicone makes it easier to reprocess the endoscope.

False! Simethicone makes it harder to reprocess the endoscope.

100

How much water should be placed in your water basin? 

Anywhere from 300 mL to 500 mL.

100

True or False: Society guidelines recommend the use of freshly prepared solution for conducting the bedside clean? 

True! You use freshly prepared enzymatic detergent! The enzymes in the sponge are protease (breaks down proteins) and amylase (breaks down carbs and starches). 

100

True or false: Exalt scopes have distal tips that must be removed after the procedure is complete.

False! The whole scope is single use.

200

This contains living and non-living microorganisms that can adhere to endoscope surfaces and may be encased in slime. 

Biofilm

200

Olympus recommends the avoidance of petroleum-based and silicone-based lubricants?

Yes. This may interfere with scope reprocessing. 

200

_____ testing is done as soon as the endoscope makes it to the reprocessing room, right before manual cleaning

Leak testing

200

Americans do what 22 times a day? 

Open the fridge! 

200

How long have duodenoscope related infections been observed (that we are aware of)?

2015 outbreaks started

300

How long after an endoscopic procedure is completed will biofilm begin to form? 

Minutes 

300

If a reprocessed scope is pulled from the cabinet and is still wet, the scope must be reprocessed again before use. 

True! 

300

_____ cleaning is the most important step of endoscope reprocessing.

Manual cleaning 

300

What does MDRO stand for and what are they?

Multi drug resistant organisms. They are antibiotic resistant bacteria. 

300

True or false: Duodenoscope infection and contamination only occurs at elevator and distal tip of the scope.

False! Heat maps have shown that infection and contamination can happen anywhere on the scope.

400

What is the longest time allowed between the Pre-Clean and the Manual Clean before additional cleaning steps need to be taken? 

1 Hour

400

The process that involves external wiping and cleaning and internal channel brushing and flushing of endoscopes

Manual Cleaning

400

Name the steps of scope reprocessing

Precleaning 

Leak testing

Manual cleaning

Rinse

High Level Disinfection

Drying

Storage

400

First step prior to handling disinfected endoscopes

Hand hygiene

400

There are two additional actions involved during pre-cleaning of double balloon endoscope, certain ultrasound endoscopes, and the latest generation duodenoscopes

Remove the balloon or disposable endcap