IP
100

on which side of the x-ray beam is penumbra (geometric un-sharpness) greater? 

on the cathode side of the beam

100
as SID increases, recorded detail ______? 

increases

100

what is the general rule in regard to imaging extremities that are casted? 

increase kVP to penetrate the cast. 

100

what determines which photo-timing cell (AEC detector) is selected? 

the part under examination. 

100

why is proper positioning important when using AEC? 

radiation exposure depends on the selected cell, requiring accurate positioning and CR alignment. 

200

what is the formula for magnification in radiography? 

image size/object size=SID/SOD

200

as SID increases, magnification_______? 

decreases

200

how much should you adjust kVp for a wet plaster cast? 

increase kVp by 8-10

200

when should the center cell be selected? 

when the CR is directed at the part (example: AP hip)

200

the receptor exposure of a radiograph produced using AEC techniques may be influenced by ________? 

adjusting receptor exposure control ( +/- 1 density)

300

How do you find SOD for the previous formula? 

SID-OID=SOD

300

what are the two types of distortion? 

1. size distortion (magnification)

2. shape distortion (elongation and foreshortening)

300

how much do you adjust your kVp to penetrate a dry plaster cast? 

increase kVp by 5-7

300

When should we not choose the center cell? 

for a PA chest exam

300

for very thing patients or pediatrics, a _______ receptor exposure control may be selected to help prevent overexposure? 

A negative receptor exposure control (-1 density)

400

EXAMPLE QUESTION: 

a radiograph image of an object measures 5 cm in diameter. If the SID is 40 inches and the OID is 5 inches, what is the actual size of the object? 

1. 5/x=40-(40-5)

2. 5/x=40/35

3. 40x=175

4. x=4.375 cm 

400

what factors affect shape distortion? 

angulation of tube/part/IR

400

how much do you adjust your kVp when penetrating a fiberglass cast? 

increase kVp by 3-4

400

why should the center cell NOT be used for a PA chest exam? 

selecting center cell will place it over the heart and may result in over-penetration of the lung fields

400

for a very large patient, a _______ receptor exposure control may be selected to help prevent under exposure? 

positive receptor exposure control

500

what does SID effect? 

recorded detail (spatial resolution) and magnification (size distortion). 

500

what factors affect size distortion? 

OID and SID

500

When is a grid required for an examination? 

when part thickness is greater than 10-13 cm

500

what cells should be selected for a PA chest exam? 

outside cells

500

how does a technologist adjust the rate of exposure when using an AEC device? 

adjust density control