MRI Safety
Parameters
Who You Gonna Call?
Coils
Scanners
600

Technologists do this to prevent RF burns on the patient's shoulder, outer arms, legs and for larger patients, their stomachs.

Place RF padding between the patient and the bore of the MRI scanner.

600

What are the two acceleration methods commonly used with MRI.

ARC and ASSET

600
...the QA fails

GE

600
This is the proper coil when scanning TMJs

Flex Coil

600

This is the reason that we do not scan whole body exams at 732 or MR2.

No coils within the scanner table

700

Where is the X/Y gradient the strongest?

at the surface of the bore

700
These four parameters determine voxel size.

FOV, frequency / phase encoding and slice thickness

700

... there is a discrepancy between the ordered exam and the patient's history.

Radiologist

700

This is one of the greatest benefits of a large coil element.

deeper penetration

~penetration = radius of coil element

700

Elastography can be successfully scanned on all scanner except...

iMRI

800

True or False: All hospital gowns are safe during an MRI scan.

False

800

You are scanning an axial image of the orbits, you see aliasing in the eyes and back of the head; your frequency direction is.

R>L

800

... there is a discrepancy between the protocol loaded on the scanner and MedWiki.

Clinical Instructor

800

These scanners utilize the dino (inkspace) coil

MR2, MR4 (with IRB), PETMR and 732

800

Fetal exams should be scanned on 3T if the gestational period is below this.

32 weeks

"32 Rule": if under 32 weeks or under 32 BMI, fetal exam should be scheduled on the 3T

900

Where is RF padding routinely placed for a MR exam under GA.

between the tips of the fingers and thighs and between the knees /calves / feet / ankles.

900

The relationship between spatial resolution and FOV is this.

Inversely related

900

... you see a strange artifact

Clinical Instructor

900

When a patient for a Ped Brain on MR2 chooses not to watch a movie, this head coil should be used.

higher channel head coil - 32ch

900

We are protocoling a large patient with a large amount of ascites. After getting a radiologist's approval, we should schedule the patient on this scanner.

MR1: 1.5T


In order to reduce the dielectric effect

1000

True or False: All Green oxygen tanks are MRI Safe

False

1000

DAILY DOUBLE: Describe five 'main' ways to increase SNR and the effect that it have on spatial resolution.


1) Increase NEX - No effect, spatial resolution

2) Increase FOV - lower spatial resolution

3) Decrease Matrix - lower spatial resolution

4) Increase Slice Thickness - lower spatial resolution

5) Increase Oversampling - No effect, spatial resolution

6) Increase Bo (more H in NMV) - No effect



1000

... you find yourself in a unique situation where you are unaware of the policy or procedure.

Management

1000

The difference in number of coils elements between the MP coil and large MP coil is.

1

1000

In a 1.5T scanner, the fat-water frequency difference will therefore be approximately this.

220 Hz


440 Hz at 3T

M