Pacemakers
SICD
Tachy
Random
Challenge Qs
100

What is the difference between Ingevity and Ingevity+?

Ingevity+

-tri-filar inner coil (Ingevity: uni-filar)

-improved terminal pin & distal coupler joints for fatigue resistance

-helix coupler extended for ext/ret., distal seals increased due to inner coil change to prevent self-extension 

-IMPROVED HELIX EXTENSION & FEWER TURNS TO EXTEND

100

How many shocks can SICD deliver?

5

100

How many max shocks are available in each therapy zone?

VT-1: 5

VT: 6

VF: 8

100

What antiobiotics elute from EluPro?

rifampin & minocycline

100

What is the Wedensky effect?

Threshold is slightly lower when decrementing down than if you were to increment up from sub-threshold 

200

What steroid is on our Ingevity+ leads?

0.91mg dexamethasone acetate


200

What is the SICD sensing floor?

80 microvolts

200

What are the QuickConvert parameters?

1 burst, 8 pulses, 88% burst cycle/coupling interval, min. interval 220ms 

output: programmable

200

The Atlas trial showed that SICD has what % fewer lead complications compared to TV-ICD?

92%

200

Define these words:

rheobase & chronaxie

rheobase (voltage): min voltage to capture tissue at infinitely long pulse width 

chronaxie: min pulse width to capture tissue at 2x rheobase voltage 

300

What amp-hour batteries are in EL Boston pacers?

1.6 amp-hours

300

What is the cathode and anode in SICD REV polarity?

can: anode

coil: cathode

300

What are SICD post shock pacing parameters?

50bpm for 30 seconds, will inhibit if senses R waves

300

What is the upper limit of vulnerability?

Max shock energy that can induce VF from R-on-T induction (vulnerable period), above which VF cannot be induced.

300

What is BiV trigger?

BiV pace triggered 8-10ms after an RV sensed event

400

What is the definition of capture?

minimum voltage and pulse width to depolarize myocardial tissue

400

What frequency does SmartPass filter?

9Hz

400

Above what HV impedance is a shock delivered monophasic?

145 ohms

400

What is ATP timeout and what happens if applied in a zone without shocks?

Period of time during which ATP is delivered for before shock is initiated. If no shocks are programmed in zone that the rhythm was detected, no shocks will be delivered 

400

What is SMART Blanking?

Shorter blanking (no noise portion), but less sensitive AGC decrement.

15ms: A blank after V sense 

37.5ms: for all other parameters 

500

What's the size difference between a Standard Longevity and EL pacer?

8mm height

500

How many Joules are delivered from a max energy SICD shock?

80 Joules


500

How many Joules are delivered from a TV-Boston ICD max energy shock?

<41 Joules

Stored energy is approx 14% higher than delivered 

500

What is the diameter of the 3501 coil?

3mm

500

What's the inner and outer diameter of the AcuityPro outer sheath?

Outer: 9.2Fr

Inner: 7.8Fr