Bushong CR
Bushong CR
Bushong CR
Bushong CR
100

what work load steps were omitted when conversion from screen film radiography to computed radiography was complete? 

The necessary reloading of the cassette. 

100

What is the approximate color of stimulating light and emitted light? 

Stimulating light: red/orange

Emitted light: blue/green

100

Why is Beam shaping required for the laser beam?  

ensures the laser beam interacts with the imaging plate. remains a circle, not distorted. 

100

how is the latent image made visible in computed radiography? 

stimulating the metastable electrons with an intense laser light beam. 

200

Identify 3 photostimulable phosphors. 

1. BaFBR:Eu

2.Bafl:Eu

3. Baff:Eu

200

What is the purpose of an optical filter positioned before the photo detector? 

To absorb scatter long wavelength laser light

200

What are the 3 subsystems of a computed radiography reader? 

1.mechanical drive system

2. optical beam shaping/collecting systems

3. computer control system 

200

What is the purpose of europium in a photostimulable phosphor? 

It is an activator. traps electrons in a metastable state. 

300

how is the latent image formed in computed radiography? 

electrons are energized by Xray interaction with excited metastable energy states, They remain in these excited states until released by exposure to laser light

300

What is the difference between fast scan and slow scan? 

Fast scan: the rapid back and fourth of the laser beam.

Slow scan: the mechanical drive of the imaging plate in through the reader. 

300

How is ghosting caused by residual laten image reduced? 

additional erasing in the CR reader.

300

What are the layer of a CR imaging plate? 

1. top (protective overcoat)

2. PSP crystal

3. r100-250 mm of phosphor

4. Reflective layer

5. base

6. antistatic layer

7.bottom (lead)

400

what causes a photostimulable phosphor to appear turbid? 

The small size of crystals, causing them to scatter incident light

400

What is the difference between an analog signal and a digital signal? 

Analog signal: continuous modulation of an electric signal

Digital Signal: Has discrete values that make it more compatible with computer techniques. 

400

What is the approximate difference in wavelength between prompt emission and stimulated emission? 

They have the same wavelength.

500

How do we reduce backscatter radiation in computed radiography? 

The imaging plate is backed by lead, which absorbs backscatter radiation. which would increase the background noise on the image. 

500

What is the difference between sampling and quantization? 

sampling: how rapidly an analog signal is interrogated. 

Quantization: After sample, amplitude of that signal is given a discrete value or quantized.

500

How should you handle a CR imaging plate?

with the same care given to the screen film cassette. 

M
e
n
u