100

This transmits visual neurological infomraiton to the brain. 

Optic Nerve

100

Digital radiographic systems are divided into these two classes.  

Direct and Indirect conversion

100

Using an 8" air gap would do this to remnant radiation exposure to a DR detector. 

Decrease it

100

This is the difference between two adjacent gray shades on a radiographic image

Radiographic contrast

100

Modern DR mobile radiographic units possess this. 

A flat panel detector for image review

200

Controls scotopic vision

Rods

200

Computers operate using this language

Binary machine

200

DR detector element saturation can result in image _____. 

Data drop

200

A radiograph with few shades of gray exhibits

High contrast

200

The primary beam exits the patient and strikes this. 

Input screen

300

The brightness of light intensity is expressed in ____. 

Lamberts

300

Analog to digital conversion involves two steps. Quantification and _____. 

Sampling

300

This is required for a fixed kVp exposure technique system. 

Optimum threshold kVp

300

With digital imaging technologies, the final brightness and contrast appearance of an image is due to____. 

Look-up table (LUT) application

300

The output screen absorbs ____ and emits _____.

Electrons; Light photons

400

The ____ the body part, the ____ the attenuation. 

Thicker; Greater

400

This controls spatial resolution

Pixel size

400

Underexposed radiographic images demonstrate this. 

Photon starvation

400

R radiographic grid with a K-factor of 1 would offer this change in image contrast. 

No change

400

The input screen absorbs ____ and emits ____.

x-ray photons and light photons

500

Bone produces less image receptor _____ because it attenuates the x-ray beam more than soft tissue does. 

Exposure

500

The fluoro halides used in CR are predominantly composed of _____. 

Bromide

500

This standard allows various vendors to share digital data without proprietary software constraints. 

Adoption of DICOM standard

500

Unacceptable levels of this cause poor spatial resolution. 

Penumbra

500

The tech will perform this trauma view of the knee to see if there is lipohemoarthrosis present. 

Cross-table lateral projection. This looks for fat-blood layering on the image. 

600

As subject density ____, image receptor exposure _____. 

Increases; Decreases

600

In CR the latent image consists of millions of ____ and _____. 

F-centers and electron holes

600

This protects a patient's medical infomraiton from unauthorized usage and access

HIPAA (Health Insurance POrtability and Accountability Act of 1996)

600

This represents a depth to the information in the digital image

Z axis

600

The shape of the image intensifier input screen helps to control this. 

Image distortion

700

Because of its tissue composition, the human breast is ____. 

A low subject contrast structure

700

Pixel bit depth determines this. 

Gray shade levels

700

The service class and object class combine to form this. 

SOP unit

700

Quantum noise impacts radiographic image quality by _____. 

Diminishing the visibility of recorded detail

700

Radiographic grids for mobile radiography should have a _______ and _______. 

Low ratio and high frequency

800

It's primary purpose is to remove a percentage of scatter photons. 

A grid

800

Indirect DR uses this as the photoconductor material. 

Amorphous silicon

800

This technology manages huge file sizes of medical images in a radiology department. 

Data compression

800

The greater the SID, the _____ the degree of magnification. 

Smaller

800

When it is determined that air-fluid level demonstration is a priority, this number of projections may be required for ______. 

Two; air-fluid levels and a normal projection of the chest

900

This is the number of grid lines per inch

Grid frequency

900

In direct digital imaging, the electronic readout of signal is accomplished by the _____. 

Thin Film Transistor (TFT)

900

This is the principal controlling factor of radiographic receptor exposure. 

mAs

900

This occurs when the tube or image receptor are not properly aligned. 

Elongation

900

This is the edge distortion problem in image intensification tubes. 

Vignetting

1000

This is intended to remove the visibility of grid lines

Grid supression software

1000

In a digital detector, the term "fill-factor" refers to the _____. 

Sensing area of the detector element

1000

The % of mAs necessary to improve the visible appearance of noise on a digital image. 

25-35%

1000

A radiographic magnification factor of -.15 indicates ______. 

Nothing since negative MF is not possible. 

1000

A fluoroscopic image that appears dark during fluoroscopy indicates this. 

Too little radiation is hitting the input phosphor