Tachy
Can you see me now?
Implant
Therapy
Follow-Up
100

Number of therapy zones that can be programmed

What is 2?

100

Sensitivity floor of the SICD

What is 80 microvolts?

100

SICD induction method

What is 50 Hz?

BONUS: What TV-ICD Induction method is this closest to?

100

Minimum shock energy of SICD

What is 80J?

100
Criteria an episode has to meet before being stored

What is device starting to charge?

200

Lowest programmable rate for a zone

What is 170 bpm?

200

SmartPass is this kind of filter at this frequency

What is a 9 Hz high pass filter?

200

Reason to flush or massage the sternal tunnel

What is to prevent air being trapped around the electrodes and causing an inappropriate shock/sensing issues?

200

Maximum numbers of shocks per episode

What is 5?

200

0% battery life means this

What is ERI or 90 days to explant?

300

These three algorithms only available in the conditional zone

What are static morphology, dynamic morphology, and QRS width?

300

Number of different sensing profiles

What is 6?

BONUS: Name them

300

Direction the can has to be placed in the pocket

What is any way the physician wants?

300

Last algorithm prior to shock delivery

What is shock confirmation?

300

Smallest amount of AF in a 24 hour period that could record an episode

What is 6 minutes?

400

Rate of post-shock pacing

What is 50 bpm?

400

These four algorithms are used in the slow rate sensing profile

What are the static template, wide complex, interval, and alternating morphology analyses?

400

Amount of time post shock pacing will last

What is 30 seconds?

BONUS: What happens to sensing?

400

Direction of the shock

What is coil to can?

400

SmartPass will turn off if either of these happens

What is 10 seconds of asystole

OR

What is 5 consecutive QRS beats <=.25mV AND 2 V-V intervals >=1400ms

500

The first N or dot within an episode will do this to the count

What is decrease it by 3?

500
Maximum number of points used to compare each beat to template

What is 41?

500

2 criteria automatic set up evaluates

What are R wave amplitudes and R:T wave ratio?

BONUS: What is the ideal minimum R:T wave ratio for the system?

500

Criteria for the end of an episode

What is 24 consecutive RR avgs at the lowest tachy zone + 40ms?

500

The AF monitor is based off these two algorithms

What are Heart Rate Density and Ventricular Scatter Algorithms?