Design Verification-1
Design Verification-2
Design Validation
Process Validation
100

Sample Size Required for Medium Risk Characteristic Attribute Data ?

29

100

How many options are listed in the procedure for Variable sampling for disposables and reusable Instruments ? 

4

100

What method is used to identify the DVAL user sample size ? 

Binomial Distribution Exact method

100

Sample Size Required for High Risk Characteristic Attribute Data ?

44

200

how do you determine the risk level for a characteristic ?

DFMEA (severity occurrence ratings) and Risk management Procedure PRO-201884 (Tri-site), PRO-202098 (Austin)

200

What are the methods used for estimating the variable data sample sizes ? 

Tolerance Interval Analysis 

K-factor analysis 

200

how many users are required for a NPD project ? 

5

200

Sample Size Required for High Risk Characteristic Variable Data ?

30

300

what is the confidence level and reliability level required for Medium risk characteristic ?

95%C, 90%R

300

How many samples are required for documentation and labeling verification ?

1

300

how many users are required for a non NPD project (design changes to existing product) ? 

User Sample Size depends if there is a impact to the User needs 

300

How do you know the variable data is normal ? 

Anderson Darling P-Value≥0.05

400

Sample Size required for Variable data Medium Risk Characteristic ? 

Depends on the Mean and Standard Deviation 

400

Sample Sizes required for Biocompatibility testing ?

dependent on the surface area of the device. refer to the Biocompatibility procedure. 

400

what is the 90% Confidence level lower bound achieved when all the 5 users have positive (Clinical acceptable) results

55%

400

What is the process capability requirement for variable data ? 

Cpk≥1 (normal data), or Ppk≥1 (non-normal data)

500

What is the method used for Attribute sample sizes ?

Bayes Success Run theorem 

(C=0 Sampling plan) 

Confidence interval formula

500

Sample Size required for IEC/EMC Testing (Disposables. Reusables and Capital Equipment)

1

500

what to do when one of the users have negative (Clinical unacceptable) results for a NPD project ? 

1) Re-evaluate the device design with 4 additional users  

2) address the root cause and re-execute Design Validation with 5 users. 

500

what do you do if Cpk or PPk < 1 ?

Implement additional controls such as 

1) increased process monitoring 

2) increased sampling 

3) gage inspection etc