Assay
System
How to handle
Other
Customer States
100

 What are some potential causes of “Pressure Errors”? Error 2008

Thick sample, Cartridge lid not closing, Incorrect sample addition

100

My module was just replaced.  Do I have to run QC?

The decision to run QC is based on your institution's laboratory guidelines.

100

Error 4016

Full system shut down

100

Where can you find the targets of each assay?

Reference Sheet (website)

Datasheet (website)

100
"I am having issues with computer freezing up and random windows updates on my system" What data would you collect to investigate this?

Log files

Windows Event logs: Admin/Security/ Set up


200

What are the targets in the Xpert MRSA NxG

SCCmec, mecA, mecC 

200

 What type of report is needed for “Hardware Errors”?

 System Log Reports

200

What are some causes of 2120 error/ 2126 error?

Gateway board

Module

200

If a customer lost 15 cartridges due to a leak and they would like replacement of cartridges.  How many kits would they be compensated for.

2 kits.  Replacements requested in whole numbers.

200

"I have an Xpress instrument and the screen says "instrument is not detected" "

Check power switches (top and bottom).  All cables are secured.  Cables are in the correct position.  Cables are not frayed.  Correct IP address.  Firewall.

300

What sample types are approved for the Xpert Xpress Strep A test?

Throat Swabs (Collection kit E swab)

300

Which instrument would you find an Automation Failure?  How do you clear an Automation Failure?

setup > change operation mode > switch to manual mode > maintenance > automation failed location > moved items from failed site to clear site > setup > change operation mode > switch to automation mode > relaunch SW

300

Request for Certificate of Analysis (COA)

Q shared drive 

Spirit ticket

300

I am getting error 5007 however I am not seeing these errors on the system log? Why?

Errors in the 5000 range will only appear on the test report.  Errors listed on the system log are typically related to the module.

300

"I ran a positive QC and got a negative result"

Confirm QC material, test procedure, get data and escalate

400

Which instruments are approved for use with the Xpert CT/NG test?

Infinity and GeneXpert

400

What type of data should we collect for a computer-related issue?

Event Log

400

A Xpress customer says their QC summary does not show their QC results, what went wrong?

They did not run their tests under the QC section but they ran it under the Home Test page

400

You have received data for another case.

Attach data to the case.  Link related case.  Case type - duplicate.

400

"I completed my Xpert Check but I don't know what to do now, how do I get my summary report if I have no internet?"

Convert GXC file to Summary Report

500

Which MRSA assay option would be utilized for pre-surgical testing

Xpert SA Nasal Complete

500

Xpert Check is means of calibrating the modules.  TRUE or FALSE

False

500

What are the parts needed for a PM?

Xpert Check kit XPERTCHECK-US-5

Software CD 950-0415

Filters 001-1537

500

Do you ERP trigger and Migrate Xpert Check case

No

Include: Xpert Kit, SW CD, and Filters.  Do not ERP trigger or Migrate.

500

"What are some maintenance tasks I can perform to ensure my GeneXpert DX system does not overheat?"

6-12 inches

Clean fans/Filters

Rotate Modules for testing