Minimum amount of insulin needed in the cartridge AFTER the tubing has been filled to complete the load process
What is 80 units
This infusion set does not require a cannula fill
What is TrusSteel
Target treatment range for Control IQ
112.5-160 mg/dl
The minimum amount of insulin needed in the cartridge AFTER the tubing has been filled to complete the load process in Tandem Mobi pump
What is 30 units
The number of personal profiles that can be programmed in a Tandem pump
What is 6
The number of air bubbles that should be removed from cartridge prior to filling with insulin
What is 2-3 big air bubbles
This infusion set comes with a 6mm cannula requires a 0.3 unit cannula fill
What is Autosoft XC or Autosoft 90
3 things does changing the correction factor affects
reduces constraints on Basal modulation, user given boluses are more impactful, automatic correction boluses are more impactful
The green pump status light indicates this on the Mobi
What is basal rate delivery
This helps optimize Dexcom/Pump connectivity
1-Wear pump with screen face out with t:slim
2-Wear pump on same side of body as Dexcom sensor with t:slim and Mobi
This alert will occur if a user taps “Bolus” but does not complete/deliver a bolus or does not use the back arrow to return to the home screen
What is the "incomplete bolus" alert
Name one benefit of tubing
Freedom to disconnect, opens up more infusion set options, wearability (short vs long with adhesive sleeve and pump clips), smaller site/less likely to rip off
Extended boluses can be set for this many hours while CIQ is on
What is 2 hours
The Tandem Mobi has an IPX rating of 28, which means it is water resistant for how many feet and hours?
What is 8 feet deep for 2 hours
The difference between an Alert and an Alarm on the pump
Alerts: notify user about things that may require additional action or acknowledgment (Low insulin, Low power, incomplete bolus, Max bolus, Incomplete load sequence)
Alarms: Notify about actual or potential stopping of insulin. Requires action to resume deliver (Resume Pump, Low Power, Empty Cartridge, Cartridge Error, Temperature, Altitude, Occlusion). Respond quickly.
If necessary, you can get the pump to register more units have been added to pump cartridge by doing this
What is going through the entire Load Process (must tap change cartridge and fill tubing)
One reason you would switch someone to the Autosoft 30 infusion set
(Name one) Bent cannulas, leaking insulin with a 90 degree set, scar tissue, improve absorption
The treatment value at which a patient would recieve an automatic correction bolus AND how often these can occur
Predicted to be over 180 mg/dl in 30 minutes
Once an hour/an hour from any previous bolus
3 ways a bolus can be initiated with the Tandem Mobi system
Bolus calculator on Mobi app
Entering override bolus in the app
The approved insulins that can be loaded into pump
What are Novolog (Aspart)/Humalog (Lispro)
After a cartridge change, the Insulin Level Indicator on the pump home screen displays an estimate of the amount of insulin in the cartridge as depicted by a + sign in front of a preset number. How many units must be delivered before the plus + sign disappears and the insulin level will display?
What is 10 units
True or false: the 5 inch tubing for Mobi is interchangeable between infusion sets
False-only comes with Autosoft XC infusion set
This happens when a patient's pump loses Dexcom connection
CIQ will stop adjusting insulin and pump will deliver insulin as programmed in personal profile, CIQ will automatically resume when pump receives Dexcom readings again.
True or false: The Tandem Mobi system requires internet connection, cell service, and bluetooth connection for use
True, but only bluetooth is required after initial setup to run the pump
Name one troubleshooting step for fixing lack of data in t:connect/Mobi apps and/or connectivity issue between phone and pump
1- Restart phone
2-Make sure app is updated/if it is, delete and re-install app
3-Unpair/Re-pair phone to pump
4-Disconnect from WiFi/or connect to WiFi (whatever is not currently the case)