Compatibility/Safety
Workflow
Trophon unit
Consumables
Log Book
100

Location for finding which probes are compatible in Trophon

100

Prior to using probe on patient, these two things should be verified.

What is sticker on clean storage bag states that the probe passed HLD and no visible bioburden on probe?

100

This sticker tells us when annual service is due.

What is PM (Preventative maintenance) sticker?

100

Regulatory verifies Trophon solution lot number and expiration date by looking at these 2 things

What is information printed on sticker and cross comparing to Trophon solution box?

100

When placing Trophon sticker in logbook, what information from previous sticker should be reviewed and why?

What is cycle number to ensure no missing cycles?

200

Meaning of the abbreviation MIFU

What is Manufacturers instructions for use?

200
Describe the types of wipes you use, in order, after removing the dirty probe cover.

What is moistened lint free cloth, sani cloth from port to base of probe, sani cloth from base of probe to tip of probe. 

200

Describe how to hang a probe in the Trophon

What is with the tip of the probe above embossed the line? 

200

Trophon solution is replaced under these 3 conditions.

What is every 30 days, when solution is empty or when solution expires?

200

What information regarding consumables is printed out on the sticker?

What is lot number and expiration date?
300

Meaning of the abbreviation SDS

What is safety data sheets

300

Route taken in Trophon room prior to HLD and then after HLD performed.

What is the clean to dirty workflow?

300

The next steps you should take if a cycle fails twice.

What is take Trophon out of service and notify leadership, re-run probe in different unit and complete PG/HRP report?

300

Steps followed when opening a new box of indicators

What is scan into Trophon(s), write date opened on box and initials of who scanned into Trophon?

300

A Trophon sticker that has passed disinfection has passed these 2 things.

What is cycle/unit passed disinfection and indicator disc passed?

400

Steps you would take to find SDS

What is The Source>Find Resources>Apps>SDS 
400

Prior to placing in Trophon, these two actions are performed.

What is visually inspect probe for bioburden and wipe with dry lint-free cloth?

400

Process used to determine/verify if an indicator disc passed disinfection

What is compare color of disc to the indicator box?

400

Best practice for storage of indicator discs

What is keep in clean, dry environment away from heat and light?

400

These are the 4 times you should start a new log page.

What is new month, new indicator, new solution, loaner unit in place?

500

Steps you would take to find MIFU online

What is The Source>Find Resources>Apps>OneSource
500

Upon removing probe from Trophon, describe 3 actions that are performed

What is inspect for bioburden, dry with lint free cloth, ensure indicator disc passed, place in clean bag, swipe badge for passed/failed cycle.

500

This feature is turned off when a probe needs HLD but does not have a Trophon tag.

What is Accutrace? 

500

Probes in biohazard bags should be secured in bag with this.

What is elastic band around bag at base of probe?
500

When starting a new log page, this is how you tell when the Trophon solution was loaded.

What is look under records on the Trophon unit menu?