Minimum fill on the t:slim
What is 95 units
This infusion set does not require a cannula fill
What is TruSteel
Target treatment range for Control IQ
112.5-160 mg/dl
One benefit of t:connect reports
1-shows basal rate modulation
2-shows automatic correction boluses AND normal food/correction boluses
3-shows time in range and basal/bolus %
4-Shows when CIQ suspended pump for lows
5-Shows time CIQ is in use
6-When Sleep/Exercise activities are being used
7- Dexcom tracings/trends
8- Pump settings including basal that is programmed vs basal that has been delivered
9- Logbook shows what basal rate was delivered by CIQ hour by hour
10- Alert/Alarm settings
Pump is watertight for this deep and this long
What is 3 feet for 30 minutes
The number of air bubbles that should be removed from cartridge prior to filling with insulin
What is 2-3 big air bubbles
What is Autosoft XC or Autosoft 90
Minimum number of hours sleep activity should run for
5 hours (should not use for naps)
People using sleep overnight on average have TIR > 80% during that time frame
Name one troubleshooting step for fixing the lack of data in t:connect and/or connectivity issue between phone and pump
1- Restart phone
2-Make sure app is updated
3-Unpair/Re-pair phone to pump
4-Disconnect from WiFi
6
If necessary, you can get the pump to register more units have been added to pump cartridge by doing this
What is going through the entire Load Process (must tap change cartridge and fill tubing)
This infusion set is manually inserted at a 30 degree angle
What is the VariSoft
Extended boluses can be set for this long while CIQ is on
2 hours
(If CIQ is off, can be set for up to 8 hours)
True or False: Alerts and Alarms showing up on t:connect app can be cleared from phone
FALSE: they must be cleared from pump
2-Wear pump on same side of body as Dexcom sensor/transmitter
What are Novolog (Aspart)/Humalog (Lispro)
One reason you would switch someone to the Autosoft 30
(Name one) Bent cannulas, leaking insulin with a 90 degree set, scar tissue/improve absorption
The treatment value at which a patient would recieve an automatic correction bolus AND how often these can occur
Predicted to be over 180 mg/dl in 30 minutes
Once an hour/an hour from any previous bolus
What you can do in mobile app if patient says they are getting annoyed by alerts/alarms showing up on phone
Turn off Pump Alerts and Alarms
AND/OR
Turn off CGM Alerts and Errors in "Settings"
One thing that can be found in current status menu
1- Last bolus
2- Current basal rate being delivered
3- Current profile settings
4- Dexcom info (i.e. when sensor was last started, how Transmitter battery is doing)
5- If patient is using mobile app/has a connection
6- What profile is active in pump
7- If CIQ is on or off
Fill estimate volume reads this many units less than it was initially filled with each new cartridge
Usually 40-60 units less until 10 units have been delivered on new cartridge; then it counts down by 5 until you get to 40 units left and then counts down by 1 unit at a time.
Process by which a patient would fix stubborn high BGs if they have already corrected with the pump
check ketones, give injection, change site regardless of when it was last changed
Check for loose connection, air in tubing, etc.
This happens when a patient's pump loses Dexcom connection
CIQ will stop adjusting insulin and pump will deliver insulin as programmed in personal profile, CIQ will automatically resume when pump receives Dexcom readings again.
If a patient has not been using their app for a long time, they should do this to ensure data uploads to the cloud prior to clinic appointments
(name one)
1-Open app 48 hours prior to appointment
2-Download pump via T:connect on computer
The difference between an Alert and an Alarm on the pump
Alerts: notify user about things that may require additional action or acknowledgment (Low insulin, Low power, incomplete bolus, Max bolus, Incomplete load sequence)
Alarms: Notify about actual or potential stopping of insulin. Requires action to resume deliver (Resume Pump, Low Power, Empty Cartridge, Cartridge Error, Temperature, Altitude, Occlusion). Respond quickly.