Vascular Access
Water
Treatment
Central Venous
Catheter
Infection
Control
100

The following are clinical presentation :

-Increased post treatment bleeding time

-Elevated venous pressures/decreased arterial pressures ; Decreased access flows; Unable to achieve prescribed blood flow rate.

- Pulsatile thrill and high pitched bruit


What is Stenosis

100

Verify test strips are not expired; Rinse Feed water sample port for 2-minutes; then rinse the sample collection cup 3 times; perform the test; interpret the test and document.

What is Total Chlorine Test

100

How much heparin be instilled in a catheter post treatment?

What is the Doctor's ordered amount?

100

This should be Easily accessible and located as close as is feasible to the immediate area or where it can be reasonably anticipated to be found; Maintain upright throughout use, replace routinely and not be allowed to overfill

What is Sharps Container

200

A complication of an AV Graft when layers  of the graft are separated during cannulation. This complication can occur from repeated cannulation in the same area of graft material.

What is Pseudoaneurysm

200

Any water system that is used to produce purified water for dialysis purposes

What is RO Purification systems

200

A catheter Arterial Pressure is greater than 250 and alarming before and after the lines have been reversed.  What should be done?

What is refer the patient to the Vascular Access (Azura) Center?

300

Death caused by Loss of blood

What is Exsanguination

300

This water is stored in a Tank; some are in the Loops

What is Product water

300

Site must be examined for signs and symptoms of Infection, dislodgement; dressing replacement must be completed prior to accessing the bloodstream

 What is Catheter Exit Site

400

Weakened areas within a blood vessel that enlarge

and contain blood. This complication can occur from trauma to the area, such as repeated cannulation, persistent hypertension, or pressure from mechanical complications such as stenosis

What is Aneurysm

400

Place RO machine in operation for a minimum of 15mins; open sample port and allow sample port to flush for 2 minutes; put on PPE; Rinse collection cup 3 times, collect sample; Dip test pad in sample for 5 seconds and remove; after 15 seconds compare test pad to color chart and document results

What is Total Hardness Testing

400

Threads and hub must be scrubbed with 70% sterile alcohol pad for 10-15 seconds and anytime caps are removed or bloodlines are disconnected.

What is Catheter Connection disinfection

500

WHO AM I……………………………

I have to be inspected

  •  listen and feel my entire length
  •  I don’t look good when am swollen.
  •  I may not do what am supposed to do when my inside is not moving well

What is a AV Fistula or AV Graft

500

Use Myron L D1 Meter and RPC strip; check Dialysate and maintain limits at +/- 0.5mS/cm of the TCD

What is conductivity 

500

How long should a Central Venous Catheter stay in a patient before it's counted against you by CMS.

What is 90 days?