.25 mm
Which dosimeter is accurate as low as 100 uGya?
film badge
According to NCRP#116, what is the embryo or fetus equivalent limit per month?
0.5 mSv
Lead aprons used in fluoroscopy must be at least?
.25 mm lead equivalent (MUST be, not SHOULD be)
TLD
According to NCRP #116, the annual occupational effective dose limit is?
50 mSv
5 minutes of elapsed fluoroscopic time
TLD
According to NCRP #116, what is the annual effective dose limit for radiography student older than age 18?
50 mSv
Dead-man switch
Which dosimeter uses aluminum oxide to record dose?
OSL
According to NCRP #116, what is the annual effective dose limit for the general public, assuming frequent exposure?
1.0 mSv
Limit use of high-level control fluoroscopy should be no more than ___ mGy per minute?
200
How often are film badges usually changed?
once a month
according to NCRP #116, what is the annual effective dose limit for the general public, assuming infrequent exposure?
5 mSv