General Information
Admit/Recovery
Medications
Procedure Room
Scopes and Disinfection
100

What is the dwell/contact time for sani wipes?

2 mins

*Wait the 2 minutes before putting the clean linens on the beds

100

How do we obtain someone's weight in admit.

Must have the patient step on the scale.

100

What is the reversal for Versed and Fentanyl?

Romazicon (flumazenil)

Initial dose: 0.2mg IV over 15 seconds

Repeat dose: Wait 45 Seconds, if no response, give 0.2mg

No more than 1mg/dose or 3mg/hr

Narcan (naloxone)

Dose: 0.4mg IV at 2-3 minute intervals

100

What do we use to suction oral secretions during a procedure?

BLUE Yankauer

** Keep in package when not using

100

What is the minimum contact/immersion time for equipment washed in enzymatic detergent solution?

2 minutes.

200

Where should the MD consent the patient?

Admit

200

Where should the charge keys be kept?

In the charge RN's pocket.

200

What is the reversal agent for agent for Propofol?

There isn't one

200

What should the official procedural timeout include?

- Patient Name & DOB

- Procedure & MD

- Allergies & Anticoagulants 

- Verify all equipment 

200

Can you run a leaking endoscope through the OER Pro?

No. It must be fully flushed during washing, and manually high level disinfected while hooked up to the MU-1 Leak Tester.

300

The surveyor asks you a question you don't know the answer to. Where can you find the answer?

- Ask a manager

- PowerDMS (all policies are here! Know your login)

- Charge binder

** Say: "I don't know the answer, but I know where to find it"

300

Once a patient is all checked into admit how do we leave them and the room?

Bed in down position, one side rail down, call light in reach, door closed and curtains half closed.

300

Medications expire after how long once drawn up?

4 hours

300

How often do the suction canisters, laundry and O2 extension tubing need to be changed?

Every case!

300

What is the maximum amount of time a scope can sit unwashed after Point of Use/Bedside cleaning?

1 Hour. After that it must be presoaked before washing.

400

How long are the BG and HCG controls good for once opened?

3 months

**BG test strips good for 4 months

400

When the patient is left in recovery after being sedated how should the bed and rails be left?

Bed in down position w/ both rails up and call light within reach.

400

If drawing up meds to be used immediately, what needs to be on the label?

- Drug Name

- Concentration 

400
You are the sedation nurse in the procedure room and the patients starts to wake up, can you suggest to the MD the dose of propofol you want to give?

No, tell the MD "The pt is beginning to arouse, would you like to re-dose?"

400

How many days maximum can Acecide-C be used before it must be changed (regardless of passing MRC Testing)?

5 Days

500

Where is the chemical spill kit located?

Under the washing sink in the back hallway

500

Where do we dispose of the used IV bags and tubing?

Red Biohazard bucket under the sink.

500

If drawing up meds ahead of time, what needs to be on the label?

- Drug name

- Concentration

- Time/ Date drawn up

- Expiration time/Date

-Initials

500

A provider asks for a 17mm dilator, what is the size of that dilator in French Units?

54 Fr

17mm x 3 = 54

500

Is it acceptable to wash 2 dilators used on separate patients at the same time, in the same sink?

No. Each dilator must be cleaned separately and per patient.