Disinfection
Parts
Menus
Online
Water
100

Cleaning the machine externally requires this solution using a damp cloth

What is 1:100 bleach solution?

100

Another name for Hansens 

What is dialyzer Connectors?

100

Choose this Menu Option to deliver a bolus or change tx Modality.

What is the Online Menu?

100

This component delivers Substitution fluid

What is the Substitution Fluid Port?

100

Place to check for residual bleach 

What is at the drain?

200

Citric Acid is Being replaced with this premixed solution

What is citrasure?

200

These Doors are removable and easily reattached.

What are EBM Doors?

200

Use this menu option to stop the blood pump to exit priming prior to connecting patient.

What is Preparation Menu?

200

This replaces Normal Saline (NS) on 5008x Machine 

What is Substitution Fluid?

200

If using a BiBag this test is no longer required for 5008X machine

What is pH?
300

This is completed every 7 days 

What is cold (bleach) disinfection?

300

Requires an exchange every 90 days OR after 100 heat disinfect OR after 13 cold disinfect

What are the Diasafe Filters

300

Choose this Menu Option for viewing the instructions for machine set-up.

What is Blood System option?

300

The amount of substitution fluid it takes for priming

What is 500ml?

300

3 Components of Dialysate 

What is Acid, Bicarb and RO (treated) water?



400

This is completed daily at end-of-shift 

What is heat disinfect?

400

This takes the place of Arterial Blood Chamber

What is the Arterial Pressure Measurement Unit?

400

Buttons that are selectable and allows changes to be made

What is blue?

400

This creates a sterile and non-pyrogenic substitution fluid.

What is the function of diasafe filter #2?

400

This is an independent test performed by the 5008x machine testing for electrical currents of dialysate.

What is Conductivity?

500

This is required when contamination is suspected with substitution port or rinse port.

What is a mandatory Disinfect?

500

Must be used with 60cc syringe to open artertial pressure measurement unit after power outage

What is the Check Valve?

500

Color of Bar during T1 Test, Preparation and Prime 

What is orange?

500

Name 5 Components of Online Plus System

What is: Substitution Fluid Pump, Rinse Port, Substitution port, Diasafe 1&2

500

Indicates Dialysate is flowing?

What are the green waves with actively moving gray bar?