This must be sprayed on all instruments prior to transporting to CIPC
Pre-Klenz
This PPE should be worn when processing instruments in the decon area
bonnet, mask, gloves, gown
They have worked at the Tinker Medical Group for 21 years
Lisa
This item should appear in every peel pack and kit that has been processed
Class 5 indicator strip
This daily test monitors the air removal efficiency of the sterilizers.
DART (DYNAMIC AIR REMOVAL TEST)
These types of instruments must remain open
Hinged (scissors, hemostats, forceps, etc)
This is the type of pressure required in the decontamination room.
Negative
They have collectively spent over 40 years in the dental career field
Lisa and Dillow
This item is required for every sterilization load and simulates the most difficult to sterilize devices ensuring that sterilization parameters reach the most challenging areas
Process Challenge Device (PCD)
This daily test verifies that the most highly resistant microorganisms are killed during sterilization.
Biological Spore test
These two visual indicators show if an item is sterilized (Wrapped kit or peel pack)
Class 5 indicator strip and outside tape or peel pack color change
These items should never be in kits when turning instruments into decon
Needle, blades, anesthetic carpules, trash
They were selected for Honor Show Choir in High School
Lisa
When taking a kit or peel pack from sterile storage always pull from the ______. (unless otherwise specified)
Right or the front
These are used to provide immediate assurance that time, temperature, and steam saturation have been met for a sterilization cycle
Class 5 chemical indicator strips
These two areas are where IFUs (Intructions for Use) can be found. (must know BOTH)
CIPC office and one source site
This daily test verifies the ultrasonic cleaner is working efficiently.
The foil test
They played in the Little League Softball World Series at age 16
Dillow
This PPE is required on the "Clean" side of CIPC
This daily test verifies the washer's ability to clean items,
All Clean test
According to the IFU, this should be used to "lubricate" the Isolite
Water
This should be done immediately entering and leaving Decon area
Hand hygiene
These 2 people have been dipped in a vat of Awesome and continuously exude awesomeness where ever they go!!
Lisa and Dillow
The three top numbers on a load label represent this. (The bottom of label is the expiration date)
1. Sterilizer used
2. Current load
3. 3 Digit Julian date
This weekly test validates the sterilizer chamber is not leaking air and is maintaining airtight integrity
Leak Test