Procedural Requirements
Random Facts You Should Know
Equipment
Operation
Calculations
100
This is the title of C-HP-1033.610
What is the Eberline Air Monitor AMS-4 Calibration and Operation
100
This is located in the Instrument Issue - bottom drawer
What is the AMS-4 response check source ?
100
Thin Window Sealed Gas Proportional
What type of detector does the AMS-4 use?
100
This is required prior to performance check and use of the AMS-4
What is a satisfactory calibration ?
100
5.33e-9uCi/mL is the product of what for what calculation?
What is the product of 1e-8uCi/mL divided by 1.875 - Alternate DAC Calculation
200
This is listed in C-HP-1033.610 Attachment 1 as - 32 to 122 degrees Fahrenheit
What is the temperature range for operation of the AMS-4?
200
This is listed in C-HP-1033.610 Att 1 as 0.3 to 4 cfm. What is the 0.3 cfm also considered?
What is the minimum flow rate - flow rate value at which a minimum flow rate failure is determined.
200
In the classic Nintendo game Super Mario Brothers, Mario hits the blocks with this body part
What is his hand?
200
This is listed on the Calibration Certificate as +/- 20% of stored efficiency
What is the Particulate Inline/Radial Efficiency Differential requirement?
200
It takes a sloth this long to travel one mile
What is a month?
300
The Xe-133 / Kr-85 DAC value is used for this setpoint?
What is the Noble Gas setpoint limit ?
300
There is enough iron in your body to make this
What is a 3 inch nail ?
300
The screen display reads "Wait for SAC / Wait for DAC". During this time the AMS-4 is performing this function.
What is updating its history and obtaining count rate and flow rate data? (page 1-6 - tech manual)
300
Might be 2121 or 0000
What are the passwords to the AMS-4?
300
This calculation compares the count rate activity in the most recent 60 sec time interval to the previous one
What is the Fast Alarm Concentration or the Net Beta Count Rate Alarm?
400
According to RP-AA-131, the radionuclide intake and CEDE dose is included in the annual and individual's annual dose record, when an individual's internal exposure exceeds 4 DAC-hours (10mRem) during a calendar year. This is the basis for which setpoint?
What is the DAC-hour setpoint?
400
The DAC value for Co-60 at 1e-8uCi/mL is the basis for our NBCR setpoint calculation because of this.
What is the most restrictive DAC value for the nuclides we expect ?
400
The AMS-4 is currently in Audible Alarm with no Visual Alarm. Explain what has happened.
What is a condition in which the AMS-4 has gone into Alarm but is now below alarm setpoint levels?
400
Coke would be this color if food coloring wasn't added
What is Green?
400
42.48LPM Air Flow Calibrator Reading calculates to what CFM?
What is 1.5 CFM?
500
6.23e10 is a factor used in calculating the NBCR setpoint. What are the units?
What are dpm mL / uCi ft3? i.e. = 2.2e6 dpm/uCi x 2.832e4 mL / ft3 C-HP-1033.610 page 18
500
This may cause increased levels of Kr-85 in the Fuel Building
What is a release of gap activity due to failed fuel pins
500
According to the tech manual, section 3-2, a soft brush and/or water rinse can be used for what AMS-4 problem
What is radioactive contamination of the detector or surrounding areas?
500
Take a Grab Air Sample, Change out the Patch, Monitor Conditions
What is the response to a CAM Alarm?
500
The instrument will count until the beta and background channel have at least 400 counts to provide a statistical accuracy of at least 10%
What is the Gamma Factor Count?