Breathing Systems
Rebreather
Non-Rebreather
Time Constants
Scavenging System
100

The APL valve must always be opened or closed? and what's a result if not done, correct?

Open, leaving it closed is FATAL!

100

Main CO2 removal method?

Chemical reaction with soda lime

100

CO2 prevention method?

High O2 flow rate (100-350 mL/kg/min)

100

Formula?

Time constant = Volume of circuit/Fresh gas flow rate 

100

Purpose of scavenging?

Remove waste gases, protect personnel.

200

Function of the breathing system?

To provide fresh gas, remove CO2 and allow for manual ventilation.

200

Minimum bag size recommendation?

3-5x tidal volume (50 mL/kg)

200

Typical use case?

Small patients <2-3kgs

200

How many T for 95% change?

3T (time constants)

200

Passive vs. active difference?

Active uses suction and passive vents naturally.

300

What is the CO2 removal type for each system?

Chemical absorbent (rebreather) and high O2 flow rate (non-rebreather)

300

Indicator soda lime needs changing?

Color change, >30 days, increased inspired CO2

300

Key advantages?

Light weight, low resistance, less dead space, expired gases completely removed 

300

You are adjusting your inhalant % on a circle system how long will it take to for it to be 95% equilibrated with the new setting, Circuit 5L, flow 1 L/min = T?

T= 5L divided by 1 L/min = 5 minutes * 3T = 15 minutes

So, at 1 L/min flow rate, your 5 L circle system will take about 15 minutes to be 95% equilibrated with the new vapor setting. 

300

Which gases do charcoal canisters NOT absorb?

Nitrous oxide

400

What happens if the one-way valve is stuck open?

Circuit becomes total dead space and CO2 rebreathing occurs

400

Average flow for maintenance?

5 mL/kg/min

400
Disadvantages?

Costly, wasteful gas use

400

Increasing flow rate does what to the time constant?

Decreases (faster equilibrium)

400

When do you replace f/air canister?

when it weighs 50g

500

What is the role of the reservoir bag and is it an area of compliance?

Gas reserve for patient and yes, it's the only area of compliance in the system

500

Benefit vs disadvantage?

Economical but more dead space 

500

Flow relation to minute volume?

2-3x MV

500

What analogy is often used with how fresh gas flow rates affects the inhaled anesthetic concentration?

Bathtube analogy 

500

OSHA's concern with chronic exposure to gases?

Reproductive & fatal risk.