1
2
3
4
5
100

In PC/AC increasing insp. pressure will do what to minute ventilation?

Increase

100

What is circ compliance?

measures compliance of circuit

100

How to view other modes on this ventilator

Mode button on the left side


Bonus: What is another way to view modes?

100

What is inspiratory rise?

How fast it will take to get to set pressure

100

Where can machine volume be found in settings?

advanced setting on insp. pressure

200

How long is the suction button activated for once the button is pressed?

2 minutes

200

Purpose of Flow Cycle %

End of breath criteria

200

T/F

The purpose of machine volume is to set a maximum vt

False

200

T/F

In inspiratory rise, 1 indicates a fast rise  in pressure and slope is steeper

True

200

Where do I go to change flow cycle % ?

The advanced setting of insp time

300

How do you get to the MIP maneuver on the ventilator screen?

Screens --- then maneuvers

300

Increasing Flow cycle % will do what to Vt and Ti

Decrease

300

T/F

When volume limit is set, and vt reaches this limit the ventilator will audibly alarm

false

300

In the monitors screen the option AutoPEEP measures what? 

Total peep

300

dAutoPEEP measures what? 

autopeep

400

When does increasing Te increase MV? Describe how this occurs.

When you have autopeep.

AutoPEEP increases total PEEP, which decreases the actual ventilating pressure.

400

What are the 3 components of an EST?

- leak test 

- circuit compliance

- o2 sensor calibration

400

Increasing machine volume from 400 ml to 600 ml will  do what to my Ti? 

no change

400

Increasing machine volume from 100 ml to 600 ml will do what to PIP?

increase

400

Machine Volume is available in which modes?

PC/AC

PC/SIMV (mandatory and assisted breaths)

500

The ventilator displays a red waveform during the inspiratory phase. Is the patient triggering breaths? 

No

mandatory breaths: Red 

Assisted/supported breaths: Yellow

500

What setting when turned on requires a double-lumen tube?

Independent lung ventilation

500

You have a COPD pt on this ventilator who has a high work of breathing. How might you specifically manipulate the rise setting to benefit this patient? 

Turn rise down

500

You perform a MIP/P100 maneuver. After you maneuver your value for P100 is -7 and MIP is 19 What do these mean? 

P100: Pt has a high work of breathing

MIP: Pt does not have adequate muscle strength, and may not be able to generate a cough

500

What does the green expiratory waveform mean?

Purging of the expiratory flow sensor