Trouble-shooting
Prescribing
Settings
Maintenace/Disinfection
Survey Readiness/Clinical Evidence
100

Check for kinks or clots in the arterial blood line. 

check the arterial access.

How to troubleshoot a Low Arterial Pressure Alarm.

100

24 hour treatment on Tablo

 XT

100

The time that blood can be stagnant before auto blood return is not allowed.

Blood clot timer

100

Explain Tablo's Heat Cycle

An hour to heat and 90-120 minutes to cool.

100

This organization oversees water regulations for dialysis.

AAMI ( Association for the Advancement of Medical Instrumentation )

200

Check that the venous needle or venous access is secure.

How to troubleshoot a Low Venous Pressure Alarm.

200

The final amount of bicarbonate that the patient will receive.

Total Buffer

200

True or False

 Smart Flow is being utilized in Acute Setting.

False

200

How often do you have to do Full Chem on Tablo and how long is the entire procedure.

Every 7 days and 6-9 hours to complete.

200

How often do we need to do water cultures and dialysate cultures?

Preconfigured devices like Tablo requires cultures to be done quarterly.

300

This alarm indicates that Tablo could not clear the air in blue line.

End treatment without returning patient's blood.

 Air in Blue line with End treatment message.

300

HD for the first 3 hrs then UF only for the last hour.

Sequential therapy.

300

Min. systolic pressure limit that would trigger a low BP alarm.

The maximum number of low BP alarm before it triggers an end of treatment. A safety feature for Home HD patients.

Systolic low alarm and BP retake limits

300

1. PDI Sani-cloth  bleach wipes

2. 1:10 dilution of household bleach.

2 cleaning agents used to disinfect Tablo

300

Per Outset how often do we need to check Chlorine and Chloramine level on Tablo?

Pre-dialysis and every 8 hours.

400

Tablo has detected that cartridge is not loaded properly, check on all 4 quadrants of the cartridge to ensure they are properly connected. If the cartridge is fully connected and the problem persists, have the user setup the txt again, checking the cartridge switches are clean. If problem persists escalate to the next tier of support.

Cartridge not Loaded.

400

 300 and 400

Maximum DFR and Qb on Tablo

400

Designed to enhance early detection of Venous needle/access dislodgement. Lowering this value makes the alarm more sensitive.

Venous Pressure Closed Limit Span

400

It's an option for those clients that require chemical disinfection for their Hep B+ patients and Hep B unknown status patients.

Quick Chem

400

What is Outset's recommendation when the URR is less than 65% or KT/V less than 1.2 

increase size of dialyzer, increase time by 15 mins.

500

Tablo has detected high conductivity 3X with no resolution and has ended the treatment. Can set right back up. 

Check if acid is connected to red and bicarb connected to blue or if two acid is used instead of one acid and one bicarb.

Check If the concentrate is correct and match prescription.

If persists escalate to next tier of support.

Unable to clear Dialysate Conductivity alarm.

500

1. IHD

2. UF only

3. SLED, PIRRT

4. Sequential

5. XT

Treatments you can prescribe on Tablo

500

Why doesn't Tablo require Conductivity Testing?

Tablo has 2 conductivity sensors.

500

Heat cycle for 1 hr followed by 30 mins quick cool using circulating water. Water required for entire cycle.

Quick Cool

500

True or False

As a TRN you can assist the staff in trouble shooting an access.

False