Check for kinks or clots in the arterial blood line.
check the arterial access.
How to troubleshoot a Low Arterial Pressure Alarm.
24 hour treatment on Tablo
XT
The time that blood can be stagnant before auto blood return is not allowed.
Blood clot timer
Explain Tablo's Heat Cycle
An hour to heat and 90-120 minutes to cool.
This organization oversees water regulations for dialysis.
AAMI ( Association for the Advancement of Medical Instrumentation )
Check that the venous needle or venous access is secure.
How to troubleshoot a Low Venous Pressure Alarm.
The final amount of bicarbonate that the patient will receive.
Total Buffer
True or False
Smart Flow is being utilized in Acute Setting.
False
How often do you have to do Full Chem on Tablo and how long is the entire procedure.
Every 7 days and 6-9 hours to complete.
How often do we need to do water cultures and dialysate cultures?
Preconfigured devices like Tablo requires cultures to be done quarterly.
This alarm indicates that Tablo could not clear the air in blue line.
End treatment without returning patient's blood.
Air in Blue line with End treatment message.
HD for the first 3 hrs then UF only for the last hour.
Sequential therapy.
Min. systolic pressure limit that would trigger a low BP alarm.
The maximum number of low BP alarm before it triggers an end of treatment. A safety feature for Home HD patients.
Systolic low alarm and BP retake limits
1. PDI Sani-cloth bleach wipes
2. 1:10 dilution of household bleach.
2 cleaning agents used to disinfect Tablo
Per Outset how often do we need to check Chlorine and Chloramine level on Tablo?
Pre-dialysis and every 8 hours.
Tablo has detected that cartridge is not loaded properly, check on all 4 quadrants of the cartridge to ensure they are properly connected. If the cartridge is fully connected and the problem persists, have the user setup the txt again, checking the cartridge switches are clean. If problem persists escalate to the next tier of support.
Cartridge not Loaded.
300 and 400
Maximum DFR and Qb on Tablo
Designed to enhance early detection of Venous needle/access dislodgement. Lowering this value makes the alarm more sensitive.
Venous Pressure Closed Limit Span
It's an option for those clients that require chemical disinfection for their Hep B+ patients and Hep B unknown status patients.
Quick Chem
What is Outset's recommendation when the URR is less than 65% or KT/V less than 1.2
increase size of dialyzer, increase time by 15 mins.
Tablo has detected high conductivity 3X with no resolution and has ended the treatment. Can set right back up.
Check if acid is connected to red and bicarb connected to blue or if two acid is used instead of one acid and one bicarb.
Check If the concentrate is correct and match prescription.
If persists escalate to next tier of support.
Unable to clear Dialysate Conductivity alarm.
1. IHD
2. UF only
3. SLED, PIRRT
4. Sequential
5. XT
Treatments you can prescribe on Tablo
Why doesn't Tablo require Conductivity Testing?
Tablo has 2 conductivity sensors.
Heat cycle for 1 hr followed by 30 mins quick cool using circulating water. Water required for entire cycle.
Quick Cool
True or False
As a TRN you can assist the staff in trouble shooting an access.
False