Environment,safety & PPE
water quality and chemistry
Manual Cleaning & Lumens
Ultra sonic & Mechanical Equipment
Verification,biofilm & Complex items
100


This type of air pressure must be maintained in the decontamination area

to prevent airborne contaminants from escaping into clean processing area


what is negative air pressure

100


This type of water is typically used for initial flushing and coarse rinsing of

soiled surgical instruments.


What is utility water?

100


To prevent generating contaminated aerosols, manual brushing of surgical

instruments must always be performed in this manner.



What is completely submerged under the water surface?


100


The microscopic implosion of sound-wave bubbles that pulls soil from

hard-to-reach instrument crevices in ultrasonic cleaners is called this.


What is cavitation?

100


A complex matrix of microorganisms that adheres tightly to instrument

surfaces and resists cleaning and disinfection is called this.


What is a biofilm?

200


This is the minimum number of air exchanges per hour required in the decontamination area.


what is 10 air exchanges per hour

200


Water that has been deionized or treated by reverse osmosis is categorized as this type of high-purity water.


what is critical water

200


This type of material should be used for brush bristles when cleaning standard surgical instruments (unless prohibited by IFU).


What is Nylon?

200



Before processing instruments in a freshly filled ultrasonic tank, this crucial preparation step must be completed to remove dissolved air.




What is degassing?


200


Biofilms begin forming on moist, untreated instrument surfaces within this time frame.


what is minutes to hours?

300


The temperature in the decontamination area should be maintained within

this specific temperature range.


what is 60-65 degrees

300


Critical water must be used during this specific rinsing phase of instrument

cleaning to prevent mineral spotting and pyrogen deposits.


what is final rinse?

300


Abrasive pads, wire brushes, and metal sponges should never be used on

stainless steel instruments because they damage this protective surface layer.



What is the passive (chromium oxide) layer?


300


Stainless steel and aluminum instruments should not be processed together in the same ultrasonic bath to prevent this type of chemical-electrical corrosion.



What is galvanic corrosion?


300


Visual inspection of cleaned instruments should always be conducted

under this type of light/magnification tool.



What is a lighted magnifying lens (or task magnifier)?


400


To prevent splash injuries from aerosolized pathogens, this piece of PPE must be worn if safety glasses alone are used without a full-face mask.


what is a full face shield

400


Water temperature used with enzymatic detergents should not exceed this threshold to prevent coagulating proteins onto instruments.


what is 140 degrees f

400


Cannulated instruments containing these hollow channels require specialized brushing and flushing protocols.



What are lumens?


400


The physical spray impact force of water against instrument surfaces

inside mechanical washers is known as this term.



What is impingement?


400


Cleaning verification tests that detect adenosine triphosphate are commonly abbreviated as this 3-letter acronym test.



What is an ATP test?


500


These specialized, heavy-duty gloves must be worn during

decontamination tasks instead of standard thin examination gloves.


what are heavy-duty utility gloves?

500


This specific enzyme in multi-enzymatic cleaners breaks down protein-

based soils like blood and tissue.


What is protease?

500


When selecting a cleaning brush for a lumen, the brush diameter must

match the lumen's inner diameter, and the brush length must be this.



What is longer than the lumen itself (so it exits the far end)?


500


In an automated washer-disinfector, high-temperature water
provides this level of microbial reduction without achieving sterilization.


What is thermal disinfection?

500


If a surgical instrument fails visual inspection or a cleaning verification test, this action must be taken immediately.



What is re-cleaning (re-processing the item through decontamination)?