Water Room
Patient Care & Machine Set Up
Infection Control
Supplies & Medication
Survey Readiness
100

What is the maximum sample volume used when performing a total chlorine test?: What is the maximum sample volume used when performing a total chlorine test?

100 mL

100

How often should vitals be monitored?

Every 30 minutes.

100

When should hand hygiene be performed during glove use?

Hand hygiene before glove application and after glove removal.

100

What information must be documented on newly opened supplies?

Date opened and initials.

100

True or False: 

Its okay to leave water room door propped?

False 

200

What is the required procedure for developing a chlorine test strip before reading the result?

One minute with a 20-second air dry. 

200

True or False: 

When a patient is prescribed a special bath, the dialysis machine must be tested with the special bath?

True

200

What should patients be washing? 

Access 

200
How long exsept good for after opening?

6 months 

200

Standard solution is good how many days after opening?

30 days 

300

How often does Vivian prefer water checks to be completed?

Every 3 hours 

300

What should be visible at all times? 

Face and access 

300

What should be offered to patients before leaving chairside?

Hand Sanitizer 

300

True or False:

Medication bottles should be labeled with date, time, and initials?

True 

300

How long are Chlorine or peroxide strips good for after opening?

3 months 

400

What actions are required if total chlorine exceeds the allowable limit?

The Medical Director, Biomed, and leadership must be notified immediately.

400
What must be engaged before testing the machine?
Air detector 
400

True or False:

Wall boxes are part of cleaning stations?

True

400

Multi-vials are good for how many days after opening?

28 days 

400

True or False: 

You can still clean the station while the patient is still in the chair? 

False 

500

What is the maximum allowable total chlorine level before treatments must be stopped?

0.1 mg/L

500

If the dialysis prescription is not being met (including dialysis flow rate or change to /inability to obtain prescribed blood flow rate) what should you do?

the reason needs to be documented and the licensed nurse informed

500

Gloves must be removed before touching this area?

Clean supply area 

500

All syringes prepared from single use vials or multidose vials should be prepared and administered within how many hours?

4

500

What can happen if you overfill the cup for chlorine check?

It can dilute the sample and lead to lower results