5008
Other Machines
RN Help
Water Treatment
Acid & Bicarb
100
High Flux
What type of dialyzers (membranes) we use now
100
Must be rinsed daily and disinfect weekly
What is WRO 300
100
Diverts dialysate directly to drain away from analyzer to avoid exposure of patient blood to unsafe dialysate due to alarms such high/low conductivity, high/low temperature.
What is the bypass function in a hemodialysis delivery system?
100
needs to be checked before tx initiation
What is total chlorine
100
What “message” on the Cecon Acid System display will indicate that the system is operational?
What is “System Active”
200
line of defense to prevent air embolism
What is air detector
200
1 Ensure conductivity and temperature are within range by confirming status box is green. (This takes approximately 5 minutes after plugging in the concentrates.) 2. Ensure dialysate flow on Home screen set at 500 mL/min. 3. Press the Test and Options Button. 4. Verify that the appropriate options have been selected i.e. dialyzer type, single needle, Pediatric. ƒ 5. Select appropriate arterial pressure alarm width. ƒ Arterial width to be set at 120 mmHg. Venous width to be set at 100 mmHg. 6. Verify: The dialysate lines are connected behind
What conditions must be met, prior to Alarm and pressure testing on the Fresenius Pediatric machine?
200
primary cause of TMP alarms
What is clothed dialyzers
200
removes chlorine and chloramine from water
What is granular activated carbon filters
200
What is the most common alarm with The CCMS system?
What is Pressure line sensor and CD reservoir
300
needs to be 85-95% full
What is arterial chamber
300
Low humidity environment Standard room temperature Range 59°-86° F Cap sealed tightly.
Where to store Test Strips?
300
Indicates that BVM is calibrated during priming.
What is the Green Funnel Icon?
300
safe level of Hydrogen Proxide
What is 0.00-0.09
300
Acid line with red label
What is K3
400
where to look first when Venous alarms are presnt
What are Transducer Protector
400
On WRO P1 < 20 psi
What Water level alarm ?
400
Double Jeopardy: Causes flow alarm
What is kinked drain line, dialysate flow not turned on or dialysate flow needs calibration.
400
RO Screen Shows: Disinfection in Progress >>>>>>
What is other RO is in disinfection
400
Must have the RN signature when performing this task.
What is when changing empty acid containers?
500
flow geometry between blood and dialysate
What is countercurrent
500
200 k light and 1 meg ohms light
What is back up system alarm limits
500
20 - 30 mins of operation max.
What is the battery capacity of running a machine during power failure?
500
90% rej rate
What is the min limit for rejection rate for RO operation?
500
CCMS display indicates suction.
What is operation mode?