What sizes are the two available focal spots on the F2 and the only one available on the SLK?
The two F2 focal spots are:
precise focus 6X28mm
extended focus 9X50mm
SLK focal spot is: 4X50mm
What are airbags used for?
The airbags displace the water to improve image quality.
Does the F2 and SLK use the same hydraulic oil?
No
F2 uses HLP32
SLK uses DE22
What are the two gap distances for inductive sensors?
What are the only sensors adjusted to 1.2mm?
1mm and 1.2mm
SCOUP and SCOUPS
11K/19/25K
What does the 230V output voltage of T1 supply?
29K?
230V supplies the PCCU and Heating Element
Heating only
What are the three focal depths of the F2 and SLK?
What is the default position of the Dual Focus PCCU.
Precise or Extended?
F2: 165mm & 180mm
SLK:150mm
Precise Focus is the default position
11K/19K/25K
What determines airbag pump run time?
29K?
Table height determines airbag pump run time.
The air compressor turns off If the pressure in the airbag is exceeding pressure in the therapy head.
How many hydraulic cylinders are on a F2?
6
Z1, Z2, X1, X2, Y, Trend.
29K
What supply voltage for the PCCU?
48V
How many switch-mode power supplies are in the 11K/19K/25K PCCU and what does it/they do?
There are two:
6.3v for heating the Thyratron Tube
24V for Switch Focus Motor and Fans
What water temp. causes 2304?
Is this code triggered from the Water Tank or Therapy Head?
After fault is resolved, how is this error condition reset?
41 degrees
It is triggered from Therapy Head.
It can be reset by the "reset safety circuit" button on the external panel.
SCOUP and SCOUPS feed information to the electronics about the state of what?
Safety Clutch
SLK
What is the supply voltage needed to operate an SLK?
115V or 230V
How many signals are designed in the SW Hand Switch and what does it/they control?
It is a double-pole switch.
1st Level: Trig1 signal Charges the Caps.
2nd Level: Trig2 signal is Trigger Enable
What are the four functions of the Water/Air Controller
1. Cushion control process
2. Degassing control process
3. Airbag control process
4. Temperature control process
Which axis is longer.... Z1 or Z2?
Which axis is longer.....X1 or X2?
Z2 Z1 is 160mm, Z2 is 180mm
X2 X1 is 160mm, X2 is 260mm
11K/19K/25
What are the three Tap Setting to connect to the mains voltages?
200VAC
220VAC
240VAC
11K/19K/25K
Where does the SWC get the energy level, trigger mode, trigger frequency, selected focal size, and enable shockwave release commands from?
The SWC gets it commands from the MPC
The Mian Process Controller receives its command from the Control Panel, then sends corresponding commands to the BMC, SWC, Water/Air Cont.
29K
If degas cycle time is longer than 10 minutes to reach "Degas Min" switch, How long is degassing switch off?
One hour until degassing is restarted.
Connect
What supply voltages are needed to operate a Connect?
X-Ray Generator - 400VAC
F2 - 208VAC
HIRIS - 208VAC
19" Rack/Remote - 208VAC