Pulse width
Dosimetry interlocks
Dosimetry Overrides
Techniques Unique to Scanning Systems
Scanning Beam Dosimetry Measurements
100

Ways to measure the amount of Beam

what is 

  • Dose
  • Charge
  • MU
100

Treatment Complete (Spot map delivered)

 what is dDone 

100

Disables interlock on max number of retries (nominally 64pC)

what is Pulse Retry Limit Error 

100

Stitching two lateral fields together to make a larger (or longer) field (e.g. for treating a spine).

What is Match Fields (“Feathering”) 

100

MLIC

multi-layer ion chamber  

What is a device for quickly making Integral Depth Dose (IDD) measurements of pristine Bragg Peaks 

200

8pC/pulse (~50 million protons)

What is the max pulse charge

200

Treatment Time T

Time in the state Beam On is less than 20% more than the prescribed time.

dTime

200

This will turn off (disable) the pulse to pulse correction in 2.4.0+.

What is Pulse Position Correction: 

200

An island (or perhaps more accurately, a donut) plan is one in which there is a region meant to receive no dose that is completely surrounded by a region of dose.

What is an Island

200

At the end of each treatment, an internal record is produced that contains all of the recorded information for that beam delivery  

What is M-DRHC (aka: Mevion (non-DICOM) Treatment Record, Dosimetry Record) 

300

 A lookup table that tells the system to fire to get the desired pulse charge

What is the pulse width curve chart

300

Beam Paused Timeout 

System has not been in the state “Beam Pause” for more than the configurable time.

What is DP max

300

Disables layer correction (no shift)

What is Layer Correction: 

300

If no beam (or very little beam) comes out, it fires the ion source again right away instead of wasting time having to come back to that spot later

What is Adding automatic retries 

300

When treatment plans are exported from the TPS to the OIS, this  intercepts the plans, breaks the spots into pulses and reorders the spots to accommodate the adaptive aperture and the range shifter’s

what is the Spot Map Converter (SMC)

400

When The Dosimetry modules samples the background for a fixed period of time to determine the mean (average) of the noise floor

What is the sound you hear when you turn the key to shoot a beam and before you push the beam enable button

400

Spot Position

Spot location (in mm, referred to isocenter) is within the configured tolerance at isocenter (TICs). Used for:  

What is dPos

400

Disables retry when pulse is below minimum charge (usually set to 0pC, effectively off)

What is Pulse Retry for Low Charge 

400

/Dosimetry.xml: InplaneSlope, CrossplaneSlope, Offset for DosX and DosY, US and DS, at each 22.5-degree gantry angle

What are Relevant XML Parameters 

400

 How the TC will recognize it as being a plan that has passed through the SMC,

What is a Tag

500

The adjustments made when a system’s Pulse Width Curve can change as the system warms and cools, so there is an automated 

What is scaling PWCs

500

Layer Shift 

Requested one-time X or Y layer shift is below configurable value.

what is dShift 

500

Disables PWC scaling – always uses loaded PWC

What is Dynamic PWC

500

You could do it with a single field, but the entrance dose (which you wish was 0%) would be 66% of the tumor dose. By dividing this into 2 uniform, parallel, opposed fields  

What is Non-Uniform Parallel Opposed 

500

Just prior to pulse delivery, the XYZ controller sends out the integrate signal for this reason

What is back ground Subtraction